HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM
Report
- Report Number
- 3011649314-2025-00978
- Event Type
- Injury
- Date Received
- August 26, 2025
- Date of Event
- April 19, 2023
- Report Date
- December 26, 2025
- Manufacturer
- PRISMATIK DENTALCRAFT, INC.
- Product Code
- DZE
- PMA / PMN Number
- K143353
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DEVICE WAS RETURNED FOR ANALYSIS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4). THIS IS ONE OF THREE COMPLAINTS FOR THE SAME PERSON. THE RELATED MDR NUMBERS ARE: 3011649314-2025-00977, 3011649314-2025-00979.
THE INVESTIGATION HAS BEEN COMPLETED, AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR HAHN TAPERED IMPLANT LOT# 6113527 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: A REVIEW OF STOCK PRODUCT WAS PERFORMED FOR HAHN TAPERED IMPLANT LOT# 6113527 AND FOUND NO ADDITIONAL PRODUCT IN STOCK. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN THE ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM (70-1154-IMP0006) USING THE RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE TREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE DESCRIPTION: "LOSS OF OSSEOINTEGRATION" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LOSS OF OSSEOINTEGRATION IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, THE PATIENT HAS A HISTORY OF BRUXISM. THE PATIENT ALSO HAS TYPE III BONE QUALITY. IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." THE MANUFACTURER'S INTERNAL REFERENCE NUMBER IS: (B)(4). . CORRECTIONS: B5: "DESCRIBE EVENT OR PROBLEM". H6: UPDATED "TYPE OF INVESTIGATION" CODE. H6: UPDATED "INVESTIGATION FINDINGS" CODE. H6: UPDATED "INVESTIGATION CONCLUSIONS" CODE.
THE PATIENT PRESENTED ON (B)(6) 2023 FOR A PRIMARY PROCEDURE ON TOOTH #7. THE PATIENT RETURNED AFTER SECOND STAGE SURGERY ON (B)(6) 2023. UPON EXAMINATION, THE PROVIDER NOTICED AN INFECTION. IT WAS DETERMINED THAT THE IMPLANT LOST OSSEOINTEGRATION. THE IMPLANT WAS REMOVED AND NOT REPLACED. IT WAS REPORTED THAT THE PATIENTS' SYMPTOMS WERE RESOLVED AFTER IMPLANT REMOVAL. NO PERMANENT INJURY WAS REPORTED. THE PATIENT'S BONE QUALITY IS TYPE III AND THE PATIENT'S ORAL HYGIENE IS REPORTED AS FAIR.
THE PATIENT PRESENTED ON (B)(6) 2023 FOR A PRIMARY PROCEDURE ON TOOTH #5, #7, AND #9. THE PATIENT RETURNED AFTER THE SECOND STAGE SURGERY ON (B)(6) 2023. UPON EXAMINATION, THE PROVIDER NOTICED AN INFECTION. IT WAS DETERMINED THAT THE IMPLANT LOST OSSEOINTEGRATION. THE IMPLANT WAS REMOVED AND NOT REPLACED. IT WAS REPORTED THAT THE PATIENTS' SYMPTOMS WERE RESOLVED AFTER IMPLANT REMOVAL. NO PERMANENT INJURY WAS REPORTED. THE PATIENT'S BONE QUALITY IS TYPE III, AND THE PATIENT'S ORAL HYGIENE IS REPORTED AS GOOD. PER THE REPORTED INFORMATION, THE PATIENT HAD A HISTORY OF BRUXISM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1466441 | HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM | HAHN TAPERED IMPLANT SYSTEM | DZE | PRISMATIK DENTALCRAFT, INC. | 70-1154-IMP0006 | 6113527 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Female | Required Intervention |