FDA Adverse Event Malfunction Summary report: N

NAVLOCK

MDR report key: 22896079 · Received August 26, 2025

Report

Report Number
1723170-2025-03073
Event Type
Malfunction
Date Received
August 26, 2025
Date of Event
June 20, 2024
Report Date
August 26, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
OLO
UDI-DI
00643169209701
PMA / PMN Number
K124004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3, H6: THE PROBE, LOT NUMBER: 230315, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE PROBE HAD A TWISTED TIP, SLIGHT GALLING AND IMPACT MARKS. THE REPORTED EVENT COULD BE DUPLICATED BY MEDTRONIC PERSONNEL. CODES B01, C07, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THORACIC PROBE TIP DEFORMITY WAS CONFIRMED DURING CLEANING AND STERILIZATION. NO PATIENT WAS PRESENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2704912 NAVLOCK ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MEDTRONIC NAVIGATION, INC. 9734680 230315 00643169209701

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown