FDA Adverse Event Malfunction Summary report: N

ENTRUST AT

MDR report key: 2289424 · Received October 12, 2011

Report

Report Number
6000144-2011-05547
Event Type
Malfunction
Date Received
October 12, 2011
Date of Event
June 8, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
Z-1172-2009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ONE POWER ON RESET (POR) FOR CRITICAL RAM PARITY ERROR, ADDR 1B52, DATA CE, OCCURRED ON (B)(6) 2011 09:41:39. ONE PATIENT ALERT FOR DEVICE CIRCUIT ERROR OCCURRED ON (B)(6) 2011 09:41:39. THE DIAGNOSTIC INFORMATION IS CONSISTENT WITH THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET. NO CHANGES WERE MADE IN PARAMETERS. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154ATG ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Other 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| 620BG27 HEART BAND