ENTRUST AT
Report
- Report Number
- 6000144-2011-05547
- Event Type
- Malfunction
- Date Received
- October 12, 2011
- Date of Event
- June 8, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S46
- Removal / Correction Number
- Z-1172-2009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ONE POWER ON RESET (POR) FOR CRITICAL RAM PARITY ERROR, ADDR 1B52, DATA CE, OCCURRED ON (B)(6) 2011 09:41:39. ONE PATIENT ALERT FOR DEVICE CIRCUIT ERROR OCCURRED ON (B)(6) 2011 09:41:39. THE DIAGNOSTIC INFORMATION IS CONSISTENT WITH THE REPORTED EVENT.
IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET. NO CHANGES WERE MADE IN PARAMETERS. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENTRUST AT | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154ATG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Other | 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| 620BG27 HEART BAND |