HAHN TAPERED IMPLANT Ø3.0 X 13 MM
Report
- Report Number
- 3011649314-2025-00957
- Event Type
- Injury
- Date Received
- August 25, 2025
- Date of Event
- April 30, 2025
- Report Date
- December 26, 2025
- Manufacturer
- PRISMATIK DENTALCRAFT, INC.
- Product Code
- DZE
- PMA / PMN Number
- K143353
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DEVICE WAS RETURNED FOR ANALYSIS HOWEVER, THE EVALUATION OF THE DEVICE IS PENDING. COMPLAINT HISTORY AND PRODUCT HISTORY RECORDS WERE REVIEWED, AND DOCUMENTATION INDICATED THE PRODUCT MET RELEASE CRITERIA. ROOT CAUSE HAS NOT BEEN IDENTIFIED. ONCE THE INVESTIGATION IS COMPLETED A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4). THIS IS ONE OF SEVEN COMPLAINTS FOR THE SAME PATIENT, RELATED MDR NUMBERS: 3011649314-2025-00920, 3011649314-2025-00954, 3011649314-2025-00955, 3011649314-2025-00956, 3011649314-2025-00958, 3011649314-2025-00959.
THE DEVICE WAS NOT RETURNED FOR ANALYSIS, A NON-VISUAL INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR HAHN TAPERED IMPLANT LOT# 6140748 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: THE STOCK PRODUCT FOR HAHN TAPERED IMPLANT LOT# 6140748 WAS REVIEWED AND FOUND TO BE IN STOCK, BUT IT IS NOT APPLICABLE FOR PHYSICAL EVALUATION SINCE THE ISSUE IS CUSTOMER RELATED. INVESTIGATION METHODS/RESULTS: THE REPORTED PRODUCT HAS NOT BEEN RETURNED TO THE COMPLAINT HANDLING TEAM TO DATE THEREFORE AN ANALYSIS OF THE PHYSICAL PRODUCT COULD NOT BE PERFORMED. ROOT CAUSE: "LACK OF PRIMARY STABILITY" AND "FAILURE TO OSSEOINTEGRATE" ARE COMMON COMPLAINTS IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY AND FAILURE TO OSSEOINTEGRATE ARE INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, THE PATIENT HAS A HISTORY OF SMOKING AND PERIODONTITIS. THE PATIENT ALSO HAS TYPE III BONE QUALITY. IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." THERE WAS ALSO AN IMPLANT FRACTURE. THE POTENTIAL ROOT CAUSE IS THE OVER-TORQUING OF THE IMPLANT DURING THE INITIAL PLACEMENT WHICH MAY HAVE CAUSED A FRACTURE. ADDITIONALLY, IT IS UNCLEAR THE METHODS OF IMPLANT PLACEMENT USED DURING THE INITIAL PROCEDURE AND THE INSERTION TORQUE VALUE. IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE IMPLANT PLACEMENT SECTION: "STEP 3: ADVANCEMENT AND FINAL SEATING - CONTINUE THREADING THE IMPLANT INTO THE OSTEOTOMY SITE USING THE PREFERRED PLACEMENT METHOD. A MINIMUM TORQUE VALUE OF 35 NCM UPON FINAL SEATING INDICATES GOOD PRIMARY STABILITY." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE PRECAUTIONS SECTION: "IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IT WAS ALSO REPORTED THAT THERE WAS AN IMPLANT AND SCREW FIT ISSUE. THE POTENTIAL CAUSE FOR THIS ISSUE MAY BE DUE TO AN OVER PREPARED OSTEOTOMY. IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE SITE PREPARATION SECTION: "IF PLACING A HAHN TAPERED IMPLANT THAT IS 3.5 MM IN DIAMETER OR GREATER, SHAPING DRILLS ARE USED SEQUENTIALLY TO WIDEN THE OSTEOTOMY TO THE MATCHING DIAMETER. TO AVOID OVER-PREPARATION, WIDENING DRILL DIAMETERS SHOULD BE USED ONLY AS NEEDED, AND IN PROPER SUCCESSION. EACH SHAPING DRILL IS LENGTH-SPECIFIC, TO MATCH THE LENGTH OF THE PRESCRIBED IMPLANT. OSTEOTOMY DEPTH MAY BE INCREASED SEQUENTIALLY, BEGINNING WITH SHORTER DRILL LENGTHS, PROVIDED SUFFICIENT DEPTH IS ACHIEVED WITH THE FINAL DRILL. SELECT THE DESIRED SHAPING DRILL, ACCOUNTING FOR BONE DENSITY AND THE SIZE OF THE IMPLANT TO BE PLACED. WITH COPIOUS IRRIGATION, DRILL TO DEPTH. THE FINAL DRILL SHOULD CORRESPOND WITH THE MATCHING IMPLANT SIZE, AS CHARTED BELOW, WITH THE GOAL OF ACHIEVING HIGH PRIMARY STABILITY UPON IMPLANT PLACEMENT." THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4). CORRECTION: D9. ADDITIONAL INFORMATION: H6 (E), (A).
A HEALTHCARE PROFESSIONAL REPORTED A HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE QUALITY IS TYPE III AND THE PATIENT'S ORAL HYGIENE IS REPORTED AS GOOD. THE PATIENT PRESENTED ON (B)(6) 2025 FOR A PRIMARY PROCEDURE ON SEVERAL TOOTH LOCATIONS #3, #5, #7, #10, #12, #14 AND #30., DURING IMPLANT PLACEMENT THE PATIENT COMPLAINT OF PAIN, UPON EXAMINATION THE PROVIDER NOTICED ABNORMAL BONE LOSS AND INFLAMMATION AROUND THE IMPLANTS AND REPORTS THE IMPLANTS "DROPPED" OR BECAME LOOSE. THE IMPLANTS ALSO FRACTURED AND THERE WAS A SCREW FITTING ISSUE AND DETERMINED THE IMPLANTS LACKED PRIMARY STABILITY AND ALSO FAILED TO OSSEOINTEGRATE, WERE REMOVED AND REPLACED. IT IS REPORTED THAT THE PATIENT SYMPTOMS RESOLVED AFTER IMPLANT WERE REMOVED AND DID NOT SUFFER HARM OR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1631170 | HAHN TAPERED IMPLANT Ø3.0 X 13 MM | HAHN TAPERED IMPLANT SYSTEM, | DZE | PRISMATIK DENTALCRAFT, INC. | 70-1154-IMP0002 | 6140748 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Male | Required Intervention |