FDA Adverse Event Malfunction Summary report: N

LIPOASPIRATE WASH SYSTEM

MDR report key: 22873151 · Received August 22, 2025

Report

Report Number
1651189-2025-09351
Event Type
Malfunction
Date Received
August 22, 2025
Date of Event
August 12, 2025
Report Date
August 12, 2025
Manufacturer
TIGER AESTHETICS MEDICAL
Product Code
QKL
PMA / PMN Number
P070004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

TIGER AESTHETICS MEDICAL COMPLAINT #: (B)(4). PATIENT AGE DEFAULTED TO "99" AS NO DATE OF BIRTH WAS PROVIDED. AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. TIGER AESTHETICS MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

TIGER AESTHETICS MEDICAL COMPLAINT #: (B)(4). DEVICE EVALUATION: G3, G6, H2, H3, H6, H10. TIGER AESTHETICS MEDICAL COMPLAINT PERFORMED AN EVALUATION USING A DIFFERENT LOT/BATCH#. THE LIKELY ROOT CAUSE OF THE VALUE STUCK ONTO THE SYRINGE IS DUE TO THE USER FORCING THE SYRINGE INTO THE EXTRACTION PORT TOO FAR. TIGER AESTHETICS MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

THE HEALTHCARE PROVIDER WAS HAVING AN ISSUE OF LOSING FAT THROUGH THE FRONT OF THE VITALITY DEVICE. UPON USING THE SYRINGE, IT WAS OBSERVED THAT THE DUCK VALVE THAT PREVENTS FAT FROM LEAKING OUT OF THE CANISTER WAS ATTACHED TO THE SYRINGE. AN ALTERNATE METHOD OF EXTRACTING THE FAT WAS USED AND MAINTAINED THE EXTRACTION POINT BEING SCREWED SHUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
345000 LIPOASPIRATE WASH SYSTEM FAT TRANSFER QKL TIGER AESTHETICS MEDICAL VIALITY-1400 A56913

Patients

Seq Age Sex Outcome Treatment
1