FDA Adverse Event Malfunction Summary report: N

NEEDLE FILTER BLUNT FILL 18X1-1/2

MDR report key: 22861860 · Received August 21, 2025

Report

Report Number
1911916-2025-00596
Event Type
Malfunction
Date Received
August 21, 2025
Date of Event
April 2, 2025
Report Date
August 28, 2025
Manufacturer
BECTON DICKINSON
Product Code
GAA
UDI-DI
00382903052110
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP: IT WAS REPORTED THE FILTER NEEDLE WAS PLACED ON THE SYRINGE AND THE MEDICATION WAS BEING WITHDRAWN, THE MEDICATION LEAKED OUT. OUR QUALITY TEAM HAS COMPLETED A DEVICE HISTORY RECORD REVIEW FOR MATERIAL NUMBER 305211 AND LOT NUMBER 4031348. THE REVIEW DID NOT IDENTIFY ANY ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE DEFECT, AND ALL QUALITY TESTS WERE WITHIN SPECIFICATION. DUE TO THE UNAVAILABILITY OF A SAMPLE FOR RETURN, A COMPREHENSIVE SAMPLE INVESTIGATION COULD NOT BE CONDUCTED. CONSEQUENTLY, THE EXACT CAUSE OF THIS INCIDENT REMAINS UNDETERMINED. SHOULD YOU ENCOUNTER ANY FURTHER ISSUES WITH OUR PRODUCT, WE WOULD APPRECIATE THE OPPORTUNITY TO PERFORM A DETAILED ANALYSIS. AT THIS TIME, NO FURTHER ACTION IS DEEMED NECESSARY.

Description of Event or Problem · 0

IT IS REPORTED LEAKAGE. EVENT DESCRIPTION: ON 09APR2025 WAS INFORMED OF AN EVENT WITH EYLEA 8MG VIAL KIT WHICH WAS CATEGORIZED WITH THE DEFECT COMPONENT ISSUE CI, LEAKER SYRINGE W/ HUB. ON (B)(6) 2025, THE OFFICE STAFF REPORTED THE FOLLOWING: ¿WHEN THE FILTER NEEDLE WAS PLACED ON THE SYRINGE AND THE MEDICATION WAS BEING WITHDRAWN, THE MEDICATION LEAKED OUT. IT LEAKED THE LUER-LOCK CONNECTION BETWEEN THE FILTER NEEDLE AND THE SYRINGE. THE DOCTOR CONFIRMED THE FILTER NEEDLE WAS ATTACHED TIGHTLY TO THE SYRINGE.¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
443320 NEEDLE FILTER BLUNT FILL 18X1-1/2 NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE GAA BECTON DICKINSON 4031348 00382903052110

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown