FDA Adverse Event Malfunction Summary report: N

DETERMINE HIV-1/2 AG/AB COMBO 25T

MDR report key: 22860738 · Received August 21, 2025

Report

Report Number
1221359-2025-00343
Event Type
Malfunction
Date Received
August 21, 2025
Date of Event
July 29, 2025
Report Date
December 16, 2025
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
MZF
UDI-DI
00811877011101
PMA / PMN Number
BP120037
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 0001011297 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 7D2648 / LOT 0001011297 AND DEVICE PART NUMBER 10732998 / LOT 101399. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 0001011297 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER, IT COULD HAVE POSSIBLY BEEN RELATED TO THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE DETERMINE HIV-1/2 AG/AB COMBO PERFORMED ON (B)(6) 2025 WITH AN UNKNOWN SAMPLE TYPE. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION INCLUDING PATIENT TREATMENT AND OUTCOME WAS PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE DETERMINE HIV-1/2 AG/AB COMBO PERFORMED ON (B)(6) 2025 WITH AN UNKNOWN SAMPLE TYPE. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION INCLUDING PATIENT TREATMENT AND OUTCOME WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1973866 DETERMINE HIV-1/2 AG/AB COMBO 25T TEST, HIV DETECTION MZF ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 0001011297 00811877011101

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown