FDA Adverse Event Malfunction Summary report: N

KONG-TL VBR SYSTEM, SCREWDRIVER FOR LOCKING SCREW

MDR report key: 22860250 · Received August 21, 2025

Report

Report Number
3006493760-2025-00004
Event Type
Malfunction
Date Received
August 21, 2025
Date of Event
July 18, 2025
Report Date
August 21, 2025
Manufacturer
ICOTEC AG
Product Code
MQP
UDI-DI
07640164847849
PMA / PMN Number
K232790
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCTION DOCUMENTATION OF THE SCREWDRIVER HAS BEEN CHECKED, NO ANOMALIES WERE DETECTED WHEN THE SCREWDRIVER WAS RELEASED TO THE MARKET. THE SCREWDRIVER HAS BEEN RETURNED FROM THE HOSPITAL. A FRACTURE PATTERN ANALYSIS HAS BEEN INITIATED. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

DURING THE LOAN CONTAINER INSPECTION AT THE MANUFACTURER IT WAS NOTICED THAT THE TIP OF A SCREWDRIVER WAS BROKEN OFF. IT IS ASSUMED THAT THE FRAGMENT HAS BROKEN OFF DURING THE LAST SURGERY. IT CAN NOT BE EXCLUDED THAT THE BROKEN OFF FRAGMENT REMAINED IN THE HEAD OF THE LOCKING SCREW AND THEREFORE IN THE PATIENT'S BODY. THE SUITABILITY OF THE MATERIAL OF THE FRAGMENT (STAINLESS STEEL 1.4112) HAS NOT BEEN EVALUATED FOR LONG-TERM TISSUE/BONE CONTACT. UNDESIRABLE LOCAL OR SYSTEMIC EFFECTS CANNOT BE EXCLUDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2410876 KONG-TL VBR SYSTEM, SCREWDRIVER FOR LOCKING SCREW KONG-TL VBR SYSTEM, SCREWDRIVER FOR LOCKING SCREW MQP ICOTEC AG 44-050 23/01 07640164847849

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown