QUICK SET
Report
- Report Number
- 3003442380-2025-13043
- Event Type
- Malfunction
- Date Received
- August 21, 2025
- Date of Event
- August 1, 2025
- Report Date
- October 7, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018051
- PMA / PMN Number
- K991759
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED ARAB EMIRATES.
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010775, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010775 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND PACKAGING IN THE MACHINE MULTIVAC 12 ON 12-DEC-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 4M00077 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 12-DEC-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 4M00078 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 12-DEC-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 4L04836 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE 04-08, ON 01-DEC-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 4L01864 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE 04-05-08, ON 17-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 4L05414 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE 05-08, ON 30-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED ARAB EMIRATES. IT WAS REPORTED THAT THE PATIENT UNABLE TO LOCK THE QUICK RELEASE AND FOUND THAT THE QUICK RELEASE WAS DAMAGED ON (B)(6) 2025. NO FURTHER INFORMATION AVAILABLE
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1973703 | QUICK SET | UNO QUICK-SET 110/9 SC1 MECA | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | MMT-396A | 6010775 | 05705244018051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |