FDA Adverse Event Malfunction Summary report: N

APEX SUPERIOR ADHESIVE SLIDE - WHITE

MDR report key: 22856311 · Received August 20, 2025

Report

Report Number
1419341-2025-00034
Event Type
Malfunction
Date Received
August 20, 2025
Date of Event
July 22, 2025
Report Date
August 21, 2025
Manufacturer
LEICA BIOSYSTEMS RICHMOND INC.
Product Code
KEW
UDI-DI
20849832023221
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER REPORTED TISSUE NO LONGER STICKS TO APEX SUPERIOR ADHESIVE SLIDE - WHITE (1440), P/N 3800080E, LOT 4900073503 IN THE WATER BATH. TRENDING ANALYSIS FROM 29 JUL 2024 TO 29 JUL 2025 REPORTED ONE (1) OTHER RELATED OCCURRENCE OF THIS ISSUE FOR THIS PRODUCT LOT. A REVIEW OF THE PRODUCT HISTORY DID NOT IDENTIFY INFORMATION IN THE MANUFACTURING RECORD FOR THIS PRODUCT THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE PROBLEM REPORTED IN THIS COMPLAINT. MANUFACTURER TESTING OF RETAIN SAMPLES WAS CONDUCTED BY CUTTING MULTIPLE MULTI-TISSUE CONTROL BLOCKS AND PERFORMING THE TISSUE ADHESION PROTOCOL FOR APEX SLIDES. AFTER MICROSCOPIC EVALUATION IT WAS FOUND THAT THERE WAS TISSUE LOSS AND TISSUE DISRUPTION. THE ROOT CAUSE FOR THE "TISSUE NO LONGER STICKS" REPORTED BY THE CUSTOMER COULD NOT BE UNEQUIVOCALLY DETERMINED FROM THE INFORMATION AVAILABLE. ALTHOUGH THE INFORMATION AVAILABLE INDICATES THAT NO ADVERSE CONSEQUENCE(S) TO A PATIENT(S) HAS BEEN REPORTED TO THE MANUFACTURER, THE CIRCUMSTANCES INVOLVED IN THIS COMPLAINT HAVE PREVIOUSLY RESULTED IN SERIOUS INJURY. IF ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

ON 22 JULY 2025, LEICA BIOSYSTEMS RECEIVED INFORMATION THAT "TISSUE NO LONGER STICKS AS ON THE SLIDE WHEN PLACING TISSUE AFTER CUTTING. USER HAS TO "FISH" FOR CUT TISSUE IN THE WATERBATH." AS OF 20 AUGUST 2025, LEICA BIOSYSTEMS HAS NOT RECEIVED INFORMATION AS TO EITHER THE FINAL STATUS OF THE PATIENT TISSUE SAMPLES INVOLVED IN THIS EVENT OR THE PATIENT IMPACT/OUTCOME. NO ADVERSE CONSEQUENCE(S) TO A PATIENT(S) HAS BEEN REPORTED TO THE MANUFACTURER TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
189716 APEX SUPERIOR ADHESIVE SLIDE - WHITE SLIDES, MICROSCOPE KEW LEICA BIOSYSTEMS RICHMOND INC. APEX SUPERIOR ADHESIVE SLIDE - WHITE (1440) 4900073503 20849832023221

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown