FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 22847088 · Received August 20, 2025

Report

Report Number
9610877-2025-52994
Event Type
Malfunction
Date Received
August 20, 2025
Date of Event
August 7, 2025
Report Date
August 20, 2025
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G4: THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EB15-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K200678. D4 UDI: "NOT DISTRIBUTED IN USA", BECAUSE PRODUCTS ARE NOT DISTRIBUTED IN USA PRODUCTS, BUT SIMILAR DEVICE PRODUCT ARE LISTED IN USA. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OPERATION CHANNEL (PRIMARY) STUCK ACCESSORY/OBJECT. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE INADEQUATE/INSUFFICIENT REPROCESSING AT THE FACILITY ON THE OPERATION CHANNEL (PRIMARY). THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT "HR-RPT-0588(CHANNEL)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. OPERATION/SUCTION CHANNEL CLOGGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1155406 PENTAX VIDEO BRONCHOSCOPE EOQ HOYA CORPORATION PENTAX TOKYO OFFICE EB11-J10

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown