FDA Adverse Event Injury Summary report: N

HAHN TAPERED IMPLANT Ø4.3 X 11.5 MM

MDR report key: 22843860 · Received August 19, 2025

Report

Report Number
3011649314-2025-00935
Event Type
Injury
Date Received
August 19, 2025
Date of Event
September 26, 2024
Report Date
January 5, 2026
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED. THE INVESTIGATION HAS BEEN COMPLETED, AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR LOT#6122036 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: THERE WAS NO STOCK PRODUCT FROM LOT#6122036 AVAILABLE FOR REVIEW. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A HAHN TAPERED IMPLANTØ4.3 X 11.5 MM (70-1154-IMP0011) USING RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED, AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE THREADING OF THE IMPLANT. (SEE ATTACHED IMAGES). THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE DESCRIPTION: "FAILURE TO OSSEOINTEGRATE" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR FAILURE TO OSSEOINTEGRATE IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE THE LOSS OF PRIMARY STABILITY AT THE OSTEOTOMY SITE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PER THE REPORTED INFORMATION, THE PATIENT HAS A HISTORY OF DIABETES AND PERIODONTITIS. IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE" IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU-570 REV 3 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

THE COMPLAINT DEVICE HAS BEEN RETURNED AND THE INVESTIGATION IS CURRENTLY ONGOING. ONCE THE RETURNED DEVICE HAS BEEN EVALUATED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT A HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE QUALITY IS TYPE II AND THEIR ORAL HYGIENE IS GOOD. THE PATIENT PRESENTED ON (B)(6) 2024 FOR A PRIMARY PROCEDURE ON TOOTH #4. ON (B)(6) 2024, AT THE SECOND STAGE SURGERY THE PATIENT PRESENTED WITH PAIN AND INFECTION. THE PROVIDER DETERMINED THAT THE IMPLANT FAILED TO OSSEOINTEGRATE AND REMOVED THE IMPLANT ON (B)(6) 2024. THE SYMPTOMS RESOLVED AFTER REMOVAL AND NO PERMANENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251339 HAHN TAPERED IMPLANT Ø4.3 X 11.5 MM HAHN TAPERED IMPLANT SYSTEM DZE PRISMATIK DENTALCRAFT, INC. 70-1154-IMP0011 6122036

Patients

Seq Age Sex Outcome Treatment
1 56 YR Female Required Intervention