FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 22829685 · Received August 18, 2025

Report

Report Number
1723170-2025-02994
Event Type
Malfunction
Date Received
August 18, 2025
Date of Event
August 1, 2025
Report Date
August 27, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
UDI-DI
00643169722187
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735821R: A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE CAMERA WAS REPLACED AND THE SYSTEM PROCEEDED TO PERFORM AS INTENDED. CODES B01, C08, AND D02 ARE APPLICABLE TO THIS ANALYSIS. THE CAMERA, LOT NUMBER: P902557, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE RETURNED PSU HAD SCRATCHES ON THE HOUSING AND LENSES. A CHECK OF THE EVENT LOG SHOWED INTERMITTENT FIRMWARE INCOMPATIBILITY DATING BACK TO FEB 2020. THERE WAS A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED. THE PSU FAILED AN ACCURACY TEST (AAK) AT .685MM WITH A PASSING THRESHOLD OF .250MM. THIS PSU HAD NOT BEEN PREVIOUSLY RETURNED FOR A FAILURE. CODES B01, C08, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

LOT NUMBER ADDED FOR CONCOMITANT PRODUCT ID: 9735821R & REMOVED FDD CODE A24 MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE SYSTEM DISPLAYED 'LOCALIZER FAULTED' AFTER THE SSD WAS REPLACED. IT WAS REPORTED THAT THE BUMP STATUS AND INTERNAL TEMPERATURE WERE FAULTED ON THE NDI TOOLBOX. IT WAS NOTED THAT THE PROBABLE CAUSE OF THE ISSUE WAS THAT THE CMOS BATTERY IN THE CAMERA HAD DIED. THERE WAS NO PATIENT PRESENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1542970 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. 9735665 00643169722187

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11...."