TECNIS IOL
Report
- Report Number
- 3012236936-2025-000218
- Event Type
- Injury
- Date Received
- August 15, 2025
- Date of Event
- July 21, 2025
- Report Date
- August 15, 2025
- Manufacturer
- AMO PUERTO RICO MFG. INC.
- Product Code
- POE
- UDI-DI
- 05050474575646
- PMA / PMN Number
- P980040
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
SECTION D6B: IF EXPLANTED, GIVE DATE: UNKNOWN, INFORMATION NOT PROVIDED. SECTION E1: TITLE, EMAIL ADDRESS, CITY, STATE, POST OFFICE OR ZIP CODE: UNKNOWN, INFORMATION NOT PROVIDED. SECTION E1: TELEPHONE NUMBER: (B)(6). SECTION H3: THE DEVICE WAS NOT RETURNED FOR EVALUATION. THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORD, COMPLAINT TRENDING WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT AFTER IMPLANTATION, THE LENS UNFOLDED SLOWLY, AND THE POSTERIOR LENS REMAINED IN THE SEMI-UNFOLDED STATE. DOCTOR HAD TO SURGICALLY INTERVENE AND ADJUST OF THE LENS POSITION. THE DOCTOR ACCIDENTALLY SCRAPED THE ANTERIOR CAPSULE AT THE OPTIC EDGE AND HAPTIC OF THE LENS, RESULTING IN THE RUPTURE OF THE ANTERIOR CAPSULE IN THE PATIENT. SUTURES WERE PERFORMED. UNDER THE MICROSCOPE, IT IS RELATIVELY CLEAR THAT THE EDGE OF OPTIC HAPTIC WAS DAMAGED. DELAY TO OVERALL PROCEDURE SURGERY TIME. IOL REMAINS IMPLANTED. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2687868 | TECNIS IOL | EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS | POE | AMO PUERTO RICO MFG. INC. | ZXR00 | 05050474575646 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |