AIA-2000
Report
- Report Number
- 3004529019-2025-00007
- Event Type
- Malfunction
- Date Received
- August 14, 2025
- Date of Event
- July 18, 2025
- Report Date
- August 14, 2025
- Manufacturer
- TOSOH HI-TEC, INC.
- Product Code
- KHO
- UDI-DI
- 04560189284616
- PMA / PMN Number
- K971103
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED ISSUE. THE FSE CONFIRMED THE REPORTED ISSUE BY REVIEW OF THE ERROR LOG. THE FSE HAD THE CUSTOMER CLEAN THE LEAK SENSOR S311 WITH COMPRESSED AIR DUE TO POSSIBLE DRIED WASH AFFECTING THE SENSOR, AND THE SENSOR IS NOW OPERATING AS EXPECTED. THE CUSTOMER ALSO NOTIFIED THE FSE THEY WERE GETTING ERROR 2239 B/F PROBE 1 PURGE FAILURE ERRORS AND NOTICED THE WASH SOLUTION WAS NOT BEING DELIVERED THROUGH B/F PROBE 1 DUE TO A DEFECTIVE SOLENOID VALVE SV290. THE FSE REPLACED THE SOLENOID VALVE SV290, THEN OBSERVED NORMAL WASH DELIVERY. THE FSE VALIDATED THE ANALYZER BY RUNNING A DAILY CHECK, QUALITY CONTROL (QC) AND PATIENT SAMPLES WITHOUT ERROR AND RESULTS WITHIN ACCEPTABLE RANGE. THE CUSTOMER FOLLOWED UP AT A LATER DATE AND NOTIFIED THE FSE THAT THE LEAK SENSOR S311 ERROR HAS RETURNED. THE FSE RETURNED TO THE CUSTOMER SITE AND CONFIRMED THE RECURRENCE. THE FSE DETERMINED THE POSSIBLE CAUSE WAS DUE TO AN INTERMITTENTLY OPERATING LEAK SENSOR S311 OR A DEFECTIVE LQD BOARD. THE FSE REPLACED THE LEAK SENSOR S311 AND THE LQD BOARD, THEN OBSERVED NORMAL LED OPERATION. THE FSE VALIDATED THE ANALYZER BY RUNNING A DAILY CHECK AND QUALITY CONTROL WITH RESULTS WITHIN ACCEPTABLE RANGE AND NO ERRORS RECURRING. NO FURTHER ACTION REQUIRED BY FIELD SERVICE. THE AIA-2000 ANALYZER IS FUNCTIONING AS EXPECTED. A 13-MONTH COMPLAINT HISTORY REVIEW AND SERVICE HISTORY REVIEW THROUGH AWARE DATE OF EVENT FOR SIMILAR COMPLAINTS WAS PERFORMED FOR SERIAL NUMBER 10640701. THERE WERE NO OTHER SIMILAR COMPLAINTS FOUND DURING THE SEARCHED PERIOD. THE AIA-2000 OPERATORS MANUAL APPENDIX 4: ERROR MESSAGES STATES THE FOLLOWING: [3005] LEAK SENSOR DETECTED LEAKAGE. CAUSE : THE LEAK SENSOR UNDER B/F UNIT DETECTED LEAKAGE. MEASUREMENT IS SUSPENDED. SOLUTION : CONTACT TOSOH SERVICE CENTER OR LOCAL REPRESENTATIVES. THE MOST PROBABLE CAUSE OF THE REPORTED ISSUE WAS DUE TO A FAILURE OF THE LEAK SENSOR S311 AND THE LQD BOARD.
A CUSTOMER REPORTED ERROR 3005 LEAK SENSOR DETECTED LEAKAGE ON THE AIA-2000 ANALYZER. THE CUSTOMER STATED THE ERROR HAS OCCURRED MULTIPLE TIMES AND HAS POWERED OFF THE ANALYZER. THE ANALYZER IS DOWN. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT WHICH RESULTED IN A DELAYED REPORTING OF PATIENT SAMPLES FOR BETA HUMAN CHORIONIC GONADOTROPIN (BHCG). THERE IS NO INDICATION OF ANY PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING OF PATIENT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 956891 | AIA-2000 | FLUOROMETER, FOR CLINICAL USE | KHO | TOSOH HI-TEC, INC. | AIA-2000 ST | N/A | 04560189284616 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |