SYRINGE 5ML SALINE FILL CHINA SP
Report
- Report Number
- 1911916-2025-00583
- Event Type
- Malfunction
- Date Received
- August 14, 2025
- Date of Event
- March 9, 2025
- Report Date
- August 26, 2025
- Manufacturer
- BECTON DICKINSON
- Product Code
- NGT
- UDI-DI
- 00382903065943
- PMA / PMN Number
- K161552
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(4). FOLLOW UP. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 306594 AND LOT NUMBER 4211514. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT SIMILAR ISSUES DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
IT IS REPORTED LEAKAGE. EVENT DESCRIPTION: ON (B)(6) 2025, WHILE SEALING THE IV LINE FOR A PATIENT AFTER INFUSION, IT WAS DISCOVERED THAT THE NEEDLE WAS LEAKING. THE NEEDLE STOPPER WAS OF POOR QUALITY, WITH EXCESS PLASTIC, MAKING IT IMPOSSIBLE TO PUSH THE SOLUTION THROUGH AND SEAL THE LINE IN A TIMELY MANNER. AFTER REPLACING THE PRE-FILLED NEEDLE, THE LINE WAS PROMPTLY SEALED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1686889 | SYRINGE 5ML SALINE FILL CHINA SP | SALINE, VASCULAR ACCESS FLUSH | NGT | BECTON DICKINSON | 4211514 | 00382903065943 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |