Report Number
3004753838-2025-223107
Event Type
Malfunction
Date Received
August 14, 2025
Date of Event
July 17, 2025
Report Date
January 30, 2026
Manufacturer
DEXCOM, INC.
Product Code
QBJ
UDI-DI
00386270002525
PMA / PMN Number
K213919
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). MFR # 3004753838-2025-223107 WAS REPORTED IN ERROR. PLEASE DISREGARD INITIAL REPORTING OF THIS EVENT AS THIS EVENT HAS NOW BEEN DEEMED NOT REPORTABLE.

Additional Manufacturer Narrative · 0

(B)(4). B5: DESCRIBE EVENT OR PROBLEM - ADDITIONAL INFORMATION. H2 TYPE OF FOLLOW UP: ADDITIONAL INFORMATION. H3: DEVICE EVALUATED BY MFG - ADDITIONAL INFORMATION. H6: ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TEST WAS PERFORMED AND FAILED. AUDIO TEST; SERIAL NUMBER VERIFICATION WAS PERFORMED AND FAILED. ALARM/ALERT MANUAL TEST WAS PERFORMED AND FAILED. SPEAKER/VIBRATOR RESISTANCE WAS PERFORMED AND PASSED. INTERNAL VISUAL INSPECTION WAS PERFORMED AND FAILED. THE PERFORMANCE DATA WAS REVIEWED FINDING ASSEMBLY ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED AS NO AUDIO OUTPUT. THE PROBABLE CAUSE WAS DETERMINED TO BE ASSEMBLY ERROR VIA PRODUCT. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
865990 DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 16909434 00386270002525

Patients

Seq Age Sex Outcome Treatment
1 70 YR Female TRESIBA