FDA Adverse Event Other Summary report: N

IVD MODEL #1136

MDR report key: 2280320 · Received September 30, 2011

Report

Report Number
MW5022488
Event Type
Other
Date Received
September 30, 2011
Date of Event
September 29, 2011
Report Date
September 30, 2011
Manufacturer
SUN NUCLEAR CORPORATION
Product Code
IYE
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

ON (B)(6) 2011, WE PERFORMED IN VIVO DOSIMETRY TO VERIFY THE SKIN DOSE FOR THE PT'S TOTAL SKIN ELECTRON BEAM TREATMENT. WE USED (B)(4) IN VIVO DOSIMETRY DEVICE (IVD) MODEL #1136, WITH 8 CHANNELS AND RF CAPABILITY. EIGHT DETECTORS, SUPPLIED WITH THE DEVICE, WERE PLACED ON THE PT'S SKIN UNDER MY SUPERVISION. THEN, THE PRESCRIBED ELECTRON BEAM TREATMENT WAS ADMINISTERED WHILE DOSE RECORDING WAS TAKING PLACE. DURING THE RADIATION TREATMENT, THE PT NOTICED PAIN AND BURNING AT THE SITES OF DETECTORS CONTACTING THE SKIN. SHE REPORTED PAIN AND BURNING AT MULTIPLE DETECTORS SITES, ALTHOUGH SHE WAS NOT CERTAIN IF ALL 8 SITES OF CONTACT OF THE DETECTORS WITH THE SKIN WERE AFFECTED. I INSPECTED THE PT IMMEDIATELY AFTER THE RADIATION TREATMENT WAS ADMINISTERED AND THE DETECTORS WERE REMOVED FROM HER SKIN. THERE WAS AN AREA OF ERYTHEMA WITH IRREGULAR PALE SKIN ELEVATION ON HER POSTERIOR NECK AT THE SITE OF THE DETECTOR #4. THIS WAS MOST CONSISTENT WITH A SUPERFICIAL ELECTRON BURN. A SIMILAR BURN, BUT WITHOUT SKIN PALLOR, WAS PRESENT ON THE ANTERIOR ABDOMEN AT THE SITE OF THE DETECTOR #5 CONTACT WITH THE SKIN. VISIBLE BURNS WERE NOT PRESENT AT THE REMAINDER OF THE DETECTOR SITES. PT REPORTED NO OTHER COMPLAINTS AND EXHIBITED NO OTHER ADVERSE SIGNS. DIGITAL PHOTOGRAPHS WERE TAKEN BY ME IMMEDIATELY AFTER THE EVENT IN ORDER TO DOCUMENT IT. I TREATED THE BURN SITES ON THE POSTERIOR NECK AND ANTERIOR ABDOMEN WITH 2% TOPICAL LIDOCAINE CREAM AND SHE WAS SENT HOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IVD MODEL #1136 IVD MODEL #1136 IYE SUN NUCLEAR CORPORATION 1136 UNK

Patients

Seq Age Sex Outcome Treatment
1 18 YR Other