SENSOR MMT-7020D1 GUARDIAN SN3 1PK OUS1
Report
- Report Number
- 2032227-2025-235241
- Event Type
- Malfunction
- Date Received
- August 13, 2025
- Date of Event
- October 28, 2023
- Report Date
- August 13, 2025
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OZP
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
THIS REPORT IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW PER CAPA 686868. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED INSIDE THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED SENSOR GLUCOSE VS. BLOOD GLUCOSE. BLOOD GLUCOSE VALUE WAS 103 MG/DL WHILE SENSOR GLUCOSE VALUE WAS 200 MG/DL. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-7020D1. THE SENSOR WAS WORN FOR 2-3 DAYS. CUSTOMER DECLINED TO REVIEW. USER STATED THEY HAD ALREADY LOST 3 SENSORS BECAUSE OF THE SG V BG ISSUES. CUSTOMER WAS ADVISED SENSOR MAY NO LONGER BE RESPONDING TO GLUCOSE CHANGES, MOST LIKELY DUE TO A SENSOR NOT SITUATED PROPERLY IN THE ISF PREVENTING SENSOR FROM PROPERLY TRACKING CHANGES IN GLUCOSE AND MAY ALSO BE RELATED TO SITE LOCATION/ROTATION, INSERTION TECHNIQUE OR TAPE TYPE/TECHNIQUE. NO PRODUCT RETURN IS EXPECTED FOR MMT-7020D1.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1392962 | SENSOR MMT-7020D1 GUARDIAN SN3 1PK OUS1 | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC MINIMED | MMT-7020D1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Unknown |