FDA Adverse Event Malfunction Summary report: N

INSYTE AUTOGUARD

MDR report key: 22798102 · Received August 13, 2025

Report

Report Number
1710034-2025-01337
Event Type
Malfunction
Date Received
August 13, 2025
Date of Event
July 28, 2025
Report Date
September 30, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903818341
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 381834 AND LOT NUMBER 5010457. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO NEW INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOGUARD RETRACTION ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: REPORTED ISSUE: INSERTION OF A VVP, THE CATHETER DID NOT RETRACT OVER THE ENTIRE NEEDLE ABOUT 1 CM NOT RETRACTED. CONSEQUENCES: NO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
563586 INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 5010457 00382903818341

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown