FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE¿

MDR report key: 22796714 · Received August 13, 2025

Report

Report Number
2024168-2025-08397
Event Type
Injury
Date Received
August 13, 2025
Date of Event
July 21, 2025
Report Date
September 19, 2025
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648013089
PMA / PMN Number
P960043
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL DEVICES REFERENCED IN B5 ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.

Additional Manufacturer Narrative · 0

VISUAL INSPECTION WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. CORRECTIONS: D4 - LOT #: UPDATED FROM 5051741 TO 5061941. D4 - EXPIRATION DATE: UPDATED FROM 3/31/2027 TO 4/30/2027. D4 - PRIMARY UDI NUMBER: UPDATED TO (B)(4). H4 - DEVICE MFG DATE: UPDATED FROM 5/17/2025 TO 6/19/2025.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A CLOSURE OF AN UNKNOWN VESSEL USING TWO PROGLIDE DEVICES RELATED TO AN UNKNOWN PROCEDURE. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED. AN UNKNOWN METHOD WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764055 PERCLOSE PROGLIDE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12673-03 5061941 08717648013089

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention