BLOOD GROUPING REAGENT ANTI-C, 5ML
Report
- Report Number
- 9610824-2025-00025
- Event Type
- Malfunction
- Date Received
- August 13, 2025
- Date of Event
- August 9, 2023
- Report Date
- August 25, 2025
- Manufacturer
- BIO-RAD MEDICAL DIAGNOSTICS GMNH
- Product Code
- QHR
- UDI-DI
- 07611969950348
- PMA / PMN Number
- 125226
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AT THE TIME WE FILED OUR INITIAL REPORT FOR THIS COMPLIANT, IT HAD NOT YET BEEN ESTABLISHED THAT THE DONOR IS NEGATIVE FOR ANTI-C ANTIGEN. SINCE WE WERE IN DOUBT IF THIS COMPLAINT MEETS FDA'S REPORTING CRITERIA, WE DECIDED TO FILE A REPORT REGARDING A POTENTIALLY FALSE NEGATIVE ANTIBODY TEST RESULT. A FALSE POSITIVE AB TEST WOULD NOT HAVE TRIGGERED A REPORT. FOR THIS REASON, OUR INITIAL REPORT WAS SUBMITTED FOR THE SERACLONE ANTI-C LOT THAT HAD ALLEGEDLY CAUSED A FALSE NEGATIVE TEST RESULT WHILE THIS FINAL REPORT IS SUBMITED FOR THE LOT THAT HAD CAUSED THE POSITIVE TEST RESULT. THIS IS OUR FINAL REPORT ON THIS INCIDENT.
THIS IS OUR INITIAL REPORT ON THIS INCIDENT.
THE CUSTOMER REPORTED DISCREPANT ANTIGEN TESTING RESULTS WITH SERACLONE ANTI-C.THE CUSTOMER REPORTED A FALSE POSITIVE TEST RESULT WITH A BLOOD UNIT FROM A DONOR KNOW AS NEGATIVE FOR ANTI-C ANTIGEN: DONOR UNIT (B)(6), NEGATIVE AND TESTED (B)(6) 2023 WITH A PREVIOUS LOT OF PRODUCTS, 9229030-01 (EXP. 13.10.2023, MANUF. 20.07.2022). DONOR UNIT (B)(6), NEGATIVE AND TESTED (B)(6) 2024 WITH THE PRODUCT LOT REPORTED, 9347050-01 (EXP. 15.02.2025, MANUF. 23.11.2023). DONOR UNIT (B)(6), POSITIVE 2+ REPEATABLE WITH ANOTHER VIAL FROM SAME PRODUCT LOT AND TESTED (B)(6) 2024 WITH THE PRODUCT LOT REPORTED, 9347050-01 (EXP. 15.02.2025, MANUF. 23.11.2023). ACCORDING TO THE LAST VERSION OF IFU, V13 APPLICABLE SINCE 30.08.2021, THE CLONE IS MS24. THE OTHER LOT 9229030-01 PREVIOUSLY USED BY THE CUSTOMER HAS BEEN MANUFACTURED 20.07.2022. THEREFORE, THESE TWO LOTS USED BY THE CUSTOMER CONTAIN THE SAME CLONE MENTIONED IN THE LAST VERSION OF IFU, THERE IS NO CHANGE REGARDING THE CLONE. ACCORDING TO THE ADDITIONAL INFORMATION, THE CUSTOMER CONFIRMED THAT THESE THREE RBC BLOOD UNITS ARE FROM THE SAME DONOR. AND THE QC TESTS ARE NOT PERFORMED WITH BIO-RAD REAGENT. IN VIEW OF THE DATE OF TESTS PERFORMED BY THE CUSTOMER AND THE FACT THAT THE REAGENTS ARE EXPIRED, NO RETENTION TESTS CAN BE PERFORMED. THE ISSUE ENCOUNTERED BY THE CUSTOMER IS MOSTLY SAMPLE RELATED WITH THE RBC BLOOD UNIT (B)(6). IT IS POSSIBLE THAT THERE IS A MISLABELING ON THE RBC DONOR BLOOD UNIT WITH THE NON-EXPECTED POSITIVE RESULT. AT THE TIME WE FILED OUR INITIAL REPORT FOR THIS COMPLIANT, IT HAD NOT YET BEEN ESTABLISHED THAT THE DONOR IS NEGATIVE FOR ANTI-C ANTIGEN. SINCE WE WERE IN DOUBT IF THIS COMPLAINT MEETS FDA'S REPORTING CRITERIA, WE DECIDED TO FILE A REPORT REGARDING A POTENTIALLY FALSE NEGATIVE ANTIBODY TEST RESULT. A FALSE POSITIVE AB TEST WOULD NOT HAVE TRIGGERED A REPORT.
THE CUSTOMER REPORTED DISCREPANT ANTIGEN TESTING RESULTS WITH SERACLONE ANTI-C. IN (B)(6) 2023, THEY TESTED A PATIENT SAMPLE FOR THE C ANTIGEN AND THE TEST RESULT WAS NEGATIVE. THE SAME PATIENT WAS TESTED AGAIN IN (B)(6) 2025 WITH ANOTHER LOT OF SERACLONE ANTI-C AND THE TEST RESULT WAS POSITIVE. THE CUSTOMER STATES THAT THERE WERE NO CHANGES REGARDING THE TEST EQUIPMENT OR PROCEDURES BETWEEN 2023 AND THE TEST IN 2025. SHE ALSO STATES TO HAVE FOLLOWED THE INSTRUCTIONS FOR USE. OUR INVESTIGATION OF THIS COMPLAINT IS STILL ONGOING. THE ORIGINAL PRODUCT THAT HAD CAUSED THE ALLEGEDLY FALSE NEGATIVE REACTION HAS ALREADY EXPIRED IN 2023.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 75868 | BLOOD GROUPING REAGENT ANTI-C, 5ML | SERACLONE ANTI-C (RH2 | QHR | BIO-RAD MEDICAL DIAGNOSTICS GMNH | 9347050-01 | 07611969950348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | SERACLONE ANTI-C, LOT 9229030-01, EXP. 2023/10/13.| SERACLONE ANTI-C, LOT 9347050-01, EXP. 02/15/2025.| SERACLONE ANTI-C, LOT 9347050-01, EXP. 2025/02/15. |