FDA Adverse Event Malfunction Summary report: N

PU-681RA

MDR report key: 22789663 · Received August 12, 2025

Report

Report Number
8030229-2025-05258
Event Type
Malfunction
Date Received
August 12, 2025
Date of Event
July 22, 2025
Report Date
October 31, 2025
Manufacturer
NIHON KOHDEN CORPORATION
Product Code
MHX
UDI-DI
04931921131640
PMA / PMN Number
K102376
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAVE BEEN GETTING BLUE SCREEN ERRORS, AND THE UNIT RESTARTS ITSELF WHICH STARTED ABOUT A WEEK AGO. TECH SUPPORT (TS) EXPLAINED THIS WOULD BE AN HDD FAILURE, BUT WE CAN DO HDD TROUBLESHOOTING TO DETERMINE IF ONLY ONE DRIVE IS BAD OR BOTH. TS EXPLAINED ALL THE STEPS TO PROPERLY TROUBLESHOOT THE HDDS. THE CUSTOMER HAS DONE THE HDD TROUBLESHOOTING, AND THE ISSUE PERSISTS REGARDLESS OF WHAT HDD HE USES. HE WOULD LIKE TO MOVE FORWARD WITH SENDING THE UNIT IN FOR A REPAIR AND GETTING REPLACEMENT DRIVES. NOT IN PATIENT USE. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAVE BEEN GETTING BLUE SCREEN ERRORS, AND THE UNIT RESTARTS ITSELF WHICH STARTED ABOUT A WEEK AGO. TECH SUPPORT (TS) EXPLAINED THIS WOULD BE AN HDD FAILURE, BUT WE CAN DO HDD TROUBLESHOOTING TO DETERMINE IF ONLY ONE DRIVE IS BAD OR BOTH. TS EXPLAINED ALL THE STEPS TO PROPERLY TROUBLESHOOT THE HDDS. THE CUSTOMER HAS DONE THE HDD TROUBLESHOOTING, AND THE ISSUE PERSISTS REGARDLESS OF WHAT HDD HE USES. HE WOULD LIKE TO MOVE FORWARD WITH SENDING THE UNIT IN FOR A REPAIR AND GETTING REPLACEMENT DRIVES. NOT IN PATIENT USE. INVESTIGATION CONCLUSION: THE CUSTOMER REPORTED RECURRING BLUE SCREEN ERRORS AND AUTOMATIC RESTARTS ON A CNS (PU-681RA). NK TECH SUPPORT COORDINATED A LOANER AND REPAIR. AT THE REPAIR CENTER, THE ISSUE WAS DUPLICATED. EVALUATION IDENTIFIED CORRUPTED SOFTWARE AND DEGRADED HARD DRIVES. BOTH HDDS WERE REPLACED AND RE-IMAGED, SYSTEM INITIALIZATION AND FUNCTIONAL CHECKS WERE COMPLETED, AND THE UNIT PASSED EXTENDED TESTING. ROOT CAUSE INVESTIGATION: THE ROOT CAUSE IS HARDWARE COMPONENT DEGRADATION OF THE HARD DRIVES WITH SOFTWARE CORRUPTION.

Description of Event or Problem · 0

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAVE BEEN GETTING BLUE SCREEN ERRORS, AND THE UNIT RESTARTS ITSELF WHICH STARTED ABOUT A WEEK AGO. TECH SUPPORT (TS) EXPLAINED THIS WOULD BE AN HDD FAILURE, BUT WE CAN DO HDD TROUBLESHOOTING TO DETERMINE IF ONLY ONE DRIVE IS BAD OR BOTH. TS EXPLAINED ALL THE STEPS TO PROPERLY TROUBLESHOOT THE HDDS. THE CUSTOMER HAS DONE THE HDD TROUBLESHOOTING, AND THE ISSUE PERSISTS REGARDLESS OF WHAT HDD HE USES. HE WOULD LIKE TO MOVE FORWARD WITH SENDING THE UNIT IN FOR A REPAIR AND GETTING REPLACEMENT DRIVES. NOT IN PATIENT USE.

Description of Event or Problem · 0

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAVE BEEN GETTING BLUE SCREEN ERRORS, AND THE UNIT RESTARTS ITSELF WHICH STARTED ABOUT A WEEK AGO. TECH SUPPORT (TS) EXPLAINED THIS WOULD BE AN HDD FAILURE, BUT WE CAN DO HDD TROUBLESHOOTING TO DETERMINE IF ONLY ONE DRIVE IS BAD OR BOTH. TS EXPLAINED ALL THE STEPS TO PROPERLY TROUBLESHOOT THE HDDS. THE CUSTOMER HAS DONE THE HDD TROUBLESHOOTING, AND THE ISSUE PERSISTS REGARDLESS OF WHAT HDD HE USES. HE WOULD LIKE TO MOVE FORWARD WITH SENDING THE UNIT IN FOR A REPAIR AND GETTING REPLACEMENT DRIVES. NOT IN PATIENT USE.