FDA Adverse Event Injury Summary report: N

HEMOSPRAY ENDOSCOPIC HEMOSTAT

MDR report key: 22786906 · Received August 12, 2025

Report

Report Number
1037905-2025-00582
Event Type
Injury
Date Received
August 12, 2025
Date of Event
July 7, 2025
Report Date
September 15, 2025
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
QAU
UDI-DI
00827002565722
PMA / PMN Number
K200972
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. A CORRECTIVE ACTION (CAPA) WAS INITIATED TO FURTHER INVESTIGATE DEVICE FAILURE DUE TO BEING UNABLE TO SPRAY POWDER. THIS DEVICE IS WITHIN THE SCOPE OF THE CAPA. PRIOR TO DISTRIBUTION, ALL HEMOSPRAY ENDOSCOPIC HEMOSTATS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION (CAPA) WAS INITIATED TO FURTHER INVESTIGATE DEVICE FAILURE DUE TO BEING UNABLE TO SPRAY POWDER. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTION. A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Additional Manufacturer Narrative · 0

INVESTIGATION EVALUATION: THE PRODUCT SAID TO BE INVOLVED WAS RETURNED IN A CLEAR PLASTIC BAG. PROVIDED WITH THE RETURN WAS AN OPEN TRAY FROM THE LOT NUMBER PROVIDED IN THE REPORT. THE LABEL MATCHES THE PRODUCT RETURNED. OUR EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT AS IT WAS DESCRIBED. NOT ALL COMPONENTS WERE INCLUDED IN THE RETURN, NO CATHETERS WERE RETURNED. THE DEVICE WAS RETURNED WITH THE ON/OFF SWITCH IN THE "OFF" POSITION. THE RED ACTIVATION KNOB WAS ENGAGED IN THE HANDLE INDICATING ACTIVATION OF THE CARBON DIOXIDE (CO2) CARTRIDGE. POWDER WAS NOT OBSERVED ON THE EXTERIOR OF THE DEVICE, WITHIN THE TRAY, NOR WITHIN THE DEVICE NOZZLE. WHEN TESTED AS RETURNED THE DEVICE DID NOT SPRAY AS INTENDED. THE CO2 CARTRIDGE DID NOT AUDIBLY DISCHARGE UPON DEACTIVATION AND THE CO2 CARTRIDGE WAS FULLY PUNCTURED. A VISUAL EXAMINATION OF THE O-RING NOTED IT HAD SHIFTED, BUT DID NOT SHOW DAMAGE FROM BEING INAPPROPRIATELY POSITIONED DURING COMPRESSION, NOR WAS IT FIXED THERE, AND WAS ABLE TO BE GENTLY MOVED BACK INTO POSITION. IT IS BELIEVED THE O-RING LIKELY SHIFTED FOLLOWING/ DURING DEACTIVATION OF THE ACTIVATION KNOB. A VISUAL EXAMINATION OF LANCE INSIDE THE HANDLE SHOWED IT TO BE POSITIONED CORRECTLY INSIDE THE HANDLE AND THE LANCE TO BE BEVELED. THE INSPECTION OF THE CO2 CARTRIDGE AND REGULATOR (LANCE AND O-RING) CONFIRM THE DEVICE WAS OF THE CURRENT DESIGN. THE FOAM WAS ORIENTED CORRECTLY WITHIN THE HANDLE. WHEN TESTED WITH A NEW CO2 CARTRIDGE AND ACTIVATION KNOB, THE DEVICE SPRAYED AS INTENDED. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: THE LABORATORY EVALUATION OF THE RETURNED DEVICE COULD NOT CONFIRM THE REPORT BECAUSE THE DEVICE SPRAYED AS INTENDED WHEN TESTED WITH A NEW CO2 CARTRIDGE AND ACTIVATION KNOB. THE ROOT CAUSE FOR THE REPORT OF UNABLE TO SPRAY IS UNKNOWN. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. CATHETER OCCLUSION CAN BE A CAUSE OF THIS DEVICE NOT BEING ABLE TO SPRAY POWDER. HOWEVER, WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE NO CATHETERS WERE RETURNED FOR EVALUATION. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. A CORRECTIVE ACTION (CAPA) WAS INITIATED TO FURTHER INVESTIGATE DEVICE FAILURE DUE TO BEING UNABLE TO SPRAY POWDER. THIS DEVICE IS WITHIN THE SCOPE OF THE CAPA. PRIOR TO DISTRIBUTION, ALL HEMOSPRAY ENDOSCOPIC HEMOSTATS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION (CAPA) WAS INITIATED TO FURTHER INVESTIGATE DEVICE FAILURE DUE TO BEING UNABLE TO SPRAY POWDER. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTION. A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 0

DURING A GASTROSCOPY WITH HEMOSTASIS FOR A MALLORY-WEISS TEAR, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE HEMOSPRAY DEVICE WAS NOT WORKING, POWDER COULD NOT BE RELEASED. [UNABLE TO SPRAY SUBJECT OF REPORT]. THE PROCEDURE WAS COMPLETED WITH ANOTHER UNKNOWN SPRAYING DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2436919 HEMOSPRAY ENDOSCOPIC HEMOSTAT HEMOSTATIC DEVICE FOR ENDOSCOPIC GASTROINTESTINAL USE QAU WILSON-COOK MEDICAL INC G56572 W4762582 00827002565722

Patients

Seq Age Sex Outcome Treatment
1 38 YR Male ENDOSCOPE - UNKNOWN MAKE OR MODEL