FDA Adverse Event Malfunction Summary report: N

GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS

MDR report key: 22786489 · Received August 12, 2025

Report

Report Number
3013164176-2025-02639
Event Type
Malfunction
Date Received
August 12, 2025
Date of Event
August 1, 2025
Report Date
August 12, 2025
Manufacturer
W. L. GORE & ASSOCIATES, INC.
Product Code
MIH
UDI-DI
00733132635337
PMA / PMN Number
P020004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

W. L. GORE & ASSOCIATES, INC. (GORE) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY GORE, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. BLANK FIELDS PRESENT ON THIS REPORT INCLUDE REQUIRED FIELDS AND FIELDS DETERMINED TO BE NOT APPLICABLE. BLANK REQUIRED FIELDS INDICATE THAT THE INFORMATION WAS NOT PROVIDED, WAS DEEMED UNAVAILABLE OR WAS NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, GORE, OR ITS ASSOCIATES THAT THE DEVICE, GORE OR ITS ASSOCIATES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

ON (B)(6) 2025, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT COMMON ILIAC ARTERY ANEURYSM AND WAS IMPLANTED WITH GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESES AND GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESES (IBE). POST DEPLOYMENT OF THE INTERNAL ILIAC ARTERY COMPONENT WHEN WITHDRAWING THE DELIVERY CATHETER, THE LEADING OLIVE AND A PORTION OF THE INNER LUMEN BROKE AND DETACHED FROM THE DELIVERY CATHETER AND REMAINED IN THE PATIENT. THE PHYSICIAN SNARED THE DETATCHED OLIVE TIP AND LUMEN AND SUCCESSFULLY WITHDREW IT WITHIN THE SHEATH. THE PATIENT TOLERATED THE PROCEDURE WITH GOOD RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2364456 GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W. L. GORE & ASSOCIATES, INC. 00733132635337

Patients

Seq Age Sex Outcome Treatment
1 78 YR Male