FDA Adverse Event Malfunction Summary report: N

IMAGIO BREAST IMAGING SYSTEM, MODEL 9100

MDR report key: 22778874 · Received August 11, 2025

Report

Report Number
3014755924-2025-04001
Event Type
Malfunction
Date Received
August 11, 2025
Date of Event
April 3, 2025
Report Date
August 11, 2025
Manufacturer
SENO MEDICAL, INC.
Product Code
QNK
UDI-DI
00864802000459
PMA / PMN Number
P200003
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

FOLLOWING LOG REVIEW AND SPEAKING TO THE SITE, THE FOLLOWING IS KNOWN CONCERNING THE ISSUE: FF2P ERROR CODE FLAGGING AT (B)(6) 2025 09:37:29, WHICH IS SOMETIMES AN EBOX SYNC TRIGGER PROBLEM, AND HAS BEEN ASSOCIATED WITH LOCKUPS NOTED IN TRAC-SSI#125. WE DID RECEIVE THE EBOX LOGS FROM THE SITE AND CONCLUDED THAT THE SYSTEM LOCKUP IS MOST LIKELY THE SAME INTERMITTENT ISSUE WE HAVE SEEN IN THIS TICKET. SEPARATELY TO THE INTERNAL LOG REVIEW, AN ADDITIONAL SUCCESSFUL EXAM WITH THREE CLIPS (SAVED) WAS PERFORMED FOLLOWING RESTART.

Description of Event or Problem · 0

DURING A PATIENT SCAN AND AFTER SAVING THE FIRST CLIP, THE SYSTEM BECAME COMPLETELY UNRESPONSIVE, I.E. NONE OF THE TOUCH TABS OR BUTTONS WERE FUNCTIONAL. AS A RESULT, SITE UNABLE TO PROCEED WITH THE SCAN AND HAD TO SHUT DOWN THE SYSTEM IN ORDER TO PREPARE IT FOR THE NEXT PATIENT. THIS WAS UNDERSTANDABLY DISTRESSING FOR THE PATIENT UNDERGOING BIOPSY AND QUITE CHALLENGING FOR OUR TEAM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2437106 IMAGIO BREAST IMAGING SYSTEM, MODEL 9100 IMAGING SYSTEM QNK SENO MEDICAL, INC. 9100 N/A 00864802000459

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown