FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ NMIC/ID-307

MDR report key: 22776481 · Received August 11, 2025

Report

Report Number
1119779-2025-02934
Event Type
Malfunction
Date Received
August 11, 2025
Date of Event
June 30, 2025
Report Date
August 28, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904492893
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR MISIDENTIFICATION OF PROTEUS AND ESCHERICHIA COLI WHEN USING PHOENIX PANEL NMIC/ID-307 (CATALOG NUMBER 449829) BATCH NUMBER UNKNOWN. CUSTOMER RETURNED PANELS, ISOLATES OR PHOENIX GENERATED LAB REPORTS WERE NOT AVAILABLE FOR THE INVESTIGATION. THE BATCH NUMBER WAS NOT PROVIDED; THEREFORE, THIS COMPLAINT IS UNCONFIRMED. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. AS NO TRENDS WERE IDENTIFIED, NO CORRECTIVE ACTIONS ARE SLATED AT THIS TIME. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.

Additional Manufacturer Narrative · 0

D4. MEDICAL DEVICE LOT #: UNKNOWN, D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA /510(K)#: G4. PMA / 510(K)# K020322, K023444, K023634, K023858, K024153, K031530, K031699, K032299, K032655, K033560, K041384, K042932, K052269, K060214, K060217, K060444, K060447, K061355, K062944, K063301, K063573, K063811, K063824, K071623, K132674, K151320, K181665, K033458 AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. H4. DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 PATIENT ISOLATES PROTEUS MIRABILIS AND ESCHERICHIA COLI ARE BEING MISIDENTIFIED. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 PATIENT ISOLATES PROTEUS MIRABILIS AND ESCHERICHIA COLI ARE BEING MISIDENTIFIED. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2003540 BD PHOENIX¿ NMIC/ID-307 SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) UNKNOWN 30382904492893

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown