FDA Adverse Event Malfunction Summary report: N

2.5MM THREE-FLUTED DRILL BIT QC/230MM/200MM CALIBRATION

MDR report key: 22771657 · Received August 11, 2025

Report

Report Number
8030965-2025-08216
Event Type
Malfunction
Date Received
August 11, 2025
Date of Event
July 25, 2025
Manufacturer
SYNTHES GMBH
Product Code
HTW
UDI-DI
07611819020252
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: D10. H3, H6: THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPEDICS; HOWEVER, PHOTOS WERE PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT ¿315.920, DRILL BIT Ø2.5 CALIBR L230/205 3FLUTE F/¿ HAS BROKEN. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCES. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE DRILL BIT Ø2.5 CALIBR L230/205 3FLUTE F/ WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, DRILL BIT Ø2.5 CALIBR L230/205 3FLUTE F/ WAS BROKEN BECAUSE OF COMPONENT FAILURE, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF MEDTECH ORTHOPEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW: PART: 315.920, SYNTHES LOT: U375319, SUPPLIER LOT: U375319, RELEASE TO WAREHOUSE DATE: 04 APR 2021, SUPPLIER: (B)(4), NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

THE DRILL BIT BROKE INTRAOPERATIVELY WHEN THE DOCTOR USED IT. ALL FRAGMENTS WERE REMOVED FROM THE PATIENT. THERE WAS NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2003907 2.5MM THREE-FLUTED DRILL BIT QC/230MM/200MM CALIBRATION BIT, DRILL HTW SYNTHES GMBH U375319 07611819020252

Patients

Seq Age Sex Outcome Treatment
1 NA Male UNK - PLATES: TRAUMA.| UNK - SCREWS: TRAUMA.