FDA Adverse Event Death Summary report: N

NAVICROSS

MDR report key: 22768337 · Received August 10, 2025

Report

Report Number
2243441-2025-00041
Event Type
Death
Date Received
August 10, 2025
Date of Event
July 10, 2025
Report Date
August 10, 2025
Manufacturer
TERUMO CORPORATION, ASHITAKA
Product Code
DQY
UDI-DI
04987892077272
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THE USER FACILITY REPORTED THAT DURING A RIGHT LOWER LIMB ANGIOGRAPHY FOR SUPERFICIAL FEMORAL ARTERY CHRONIC TOTAL OCCLUSION, ACCESS WAS OBTAINED VIA THE LEFT COMMON FEMORAL ARTERY USING A SIX FRENCH COOK ANSEL SHEATH. A 90-CENTIMETER QUICK-CROSS CATHETER AND A 260-CENTIMETER ANGLED GLIDEWIRE WERE USED TO CROSS THE CHRONIC TOTAL OCCLUSION, BUT THE ATTEMPT WAS UNSUCCESSFUL. SUBSEQUENTLY, ACCESS WAS OBTAINED VIA THE RIGHT ANTERIOR TIBIAL ARTERY USING AN 0.018-INCH, 300-CENTIMETER TRACK WIRE AND A 0.018-INCH, 135-CENTIMETER NAVICROSS CATHETER. THE NAVICROSS CATHETER WAS ADVANCED THROUGH THE ANTERIOR TIBIAL ARTERY, POPLITEAL ARTERY, AND INTO THE CHRONIC TOTAL OCCLUSION CAP. THE TRACK WIRE AND NAVICROSS CATHETER WERE ADVANCED INTO THE SIX FRENCH ANSEL SHEATH. THE WIRE WAS FLOSSED THROUGH THE SHEATH, WITH THE TRACK WIRE EXTENDING THROUGH THE SYSTEM. THE PHYSICIAN ATTEMPTED TO WITHDRAW THE NAVICROSS CATHETER VIA THE PEDAL ACCESS. DURING REMOVAL, THE NAVICROSS CATHETER DID NOT ADVANCE OUT AS EXPECTED. ONE PHYSICIAN HELD THE WIRE AT THE GROIN SHEATH, WHILE A FELLOW MANAGED THE PEDAL WIRE, AND ANOTHER PHYSICIAN ATTEMPTED TO PULL THE NAVICROSS CATHETER. DURING THIS PROCESS, THE CATHETER SEPARATED. THE BROKEN PORTION WAS REMOVED, BUT ANOTHER SEGMENT REMAINED ATTACHED TO THE TRACK WIRE. THE TRACK WIRE WAS CUT TO FACILITATE REMOVAL OF THE REMAINING CATHETER SEGMENT. THE BROKEN NAVICROSS CATHETER AND THE SEGMENT ATTACHED TO THE WIRE WERE RETAINED. THE DISTAL END OF THE TRACK WIRE WAS DISCARDED. ALL COMPONENTS OF THE NAVICROSS CATHETER AND TRACK WIRE WERE SUCCESSFULLY REMOVED FROM THE PATIENT. A SHEATH WAS PLACED IN THE PEDAL ACCESS. A QUICK-CROSS CATHETER WAS USED FROM THE GROIN TO ACCESS THE TIBIAL ARTERY. A SPARTACORE WIRE WAS ADVANCED DOWN THE LEG, AND A HAWK DEVICE WAS USED FOR ATHERECTOMY. AFTER SEVERAL PASSES WITH THE HAWK DEVICE, THE PATIENT DEVELOPED TACHYCARDIA AND HYPOTENSION. ANGIOGRAPHIC IMAGING REVEALED EXTRAVASATION IN THE ILIAC ARTERY AND AORTA. A CODA BALLOON WAS DEPLOYED, AND AORTIC/ILIAC STENTS WERE PLACED. DESPITE THESE INTERVENTIONS, THE PATIENT EXPIRED. THE PROCEDURE PERFORMED WAS A LEFT LEG ANGIOGRAPHY. THE EVENT OCCURRED INTRA-OPERATIVELY. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250 CCS. ADDITIONAL INFORMATION WAS RECEIVED ON 14 JUL 2025: THE SAMPLE SHOWED BLOOD AND FLUID CONTAMINATION BUT NO PRESENCE OF ANTI-CANCER AGENTS. A PHYSICIAN INVOLVED IN THE CASE STATED THAT THERE WAS NO INDICATION THAT TERUMO PRODUCTS CAUSED OR CONTRIBUTED TO THE PATIENT'S DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2003709 NAVICROSS CATHETER, PERCUTANEOUS DQY TERUMO CORPORATION, ASHITAKA NC18131 250307 04987892077272

Patients

Seq Age Sex Outcome Treatment
1 82 YR Female Death 6F COOK ANSEL SHEATH| 90CM QUICKCROSS WITH A 260CM ANGLE GLIDEWIRE.