FDA Adverse Event Other Summary report: N

DURA-STICK II SELF-ADHESIVE ELECTRODES

MDR report key: 2275494 · Received October 3, 2011

Report

Report Number
1022819-2011-00011
Event Type
Other
Date Received
October 3, 2011
Date of Event
August 1, 2011
Report Date
September 30, 2011
Manufacturer
DJO LLC
Product Code
IPF
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADD'L DATA: ACTUAL ELECTRODES USED DURING THE EVENT WERE NOT RETURNED FOR EVAL. UNUSED ELECTRODES FROM SAME PACKAGE WERE RETURNED AND EVALUATED WITH NO ISSUES FOUND. CORRECTED DATA: REPORT ORIGINALLY ISSUED AND SUBMITTED WITH INCOMPLETE MFR REPORT NUMBER 1022819-2011-00009. MFR REPORT NUMBER CORRECTED TO 1022819-2011-00011 AS SHOWN.

Description of Event or Problem · 1

THE PT CLAIMED THEY WERE BURNED WHEN USING DJO SUPPLIED ELECTRODES WITH A NON DJO ELECTROSTIMULATOR. THE SERIOUSNESS OF THE BURN WAS NOT PROVIDED AND NO MEDICAL INTERVENTION WAS REQUIRED. END USER WAS USING DJO DURA-STICK II ELECTRODES WITH A NON DJO ELECTROSTIMULATOR DEVICE. AS THIS DEVICE WAS NOT RETURNED TO DJO THE FUNCTIONAL CONDITION IS UNK. UNUSED ELECTRODES WERE RETURNED TO DJO AND EVALUATED AND FOUND TO BE FULLY OPERATIONAL. AS DJO CANNOT CONCLUSIVELY DETERMINE IF THE CAUSE OF THE REPORTED EVENT IS DUE TO THE NON DJO ELECTROSTIMULATOR THIS EVENT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURA-STICK II SELF-ADHESIVE ELECTRODES ELECTRODES IPF DJO LLC 4202 915612

Patients

Seq Age Sex Outcome Treatment
1 Other