FDA Adverse Event Malfunction Summary report: N

HANDICARE ACCESSIBILITY LTD.

MDR report key: 22752524 · Received August 8, 2025

Report

Report Number
3011268530-2025-00003
Event Type
Malfunction
Date Received
August 8, 2025
Date of Event
June 23, 2025
Report Date
August 5, 2025
Manufacturer
HANDICARE ACCESIBILITY LTD.
Product Code
PCD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

RAIL FOUND TO HAVE SHEARED AT WELD JOIN. MINIMAL CHANCE OF FAILURE OCCURRING WITH PREVIOUS WELDING ((B)(4)), HOWEVER CHANGES TO PROCESS BETWEEN TIG & MIG WELDING OCCURRED AS AN IMPROVEMENT IN 2022. IMPLEMENTATION OF LOAD TESTING AT START AND END OF SHIFT FOR VALIDATION OF WELD PROCESS.

Description of Event or Problem · 0

TECH REPORTED THAT THE END USER TRIPPED OUT OF THE LIFT. HOWEVER, THEY WERE NOT INJURED AS SUPPORTED THEMSELVES AGAINST THE WALL. TECH COMMENTED THAT A LIFT RAIL CONNECTION APPEARED OPEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223487 HANDICARE ACCESSIBILITY LTD. STAIRLIFT PCD HANDICARE ACCESIBILITY LTD. 2000

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown