QUICK-SET
Report
- Report Number
- 3003442380-2025-12564
- Event Type
- Injury
- Date Received
- August 8, 2025
- Date of Event
- July 20, 2025
- Report Date
- August 11, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- PMA / PMN Number
- K991759
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED DATE OF EVENT UNDER B3. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY - THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6001937 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. TEST RESULTS: COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001937 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 77 PACKAGING IN THE MULTIVAC12, ON29/JUN/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 11/AUG/2025 AGAINST HARM CODE UNTREATED DIABETIC KETOACIDOSIS WHICH THE PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY A HEALTH CARE PROVIDER (HCP) OR REQUIRES EMERGENCY ADVANCED LIFE SUPPORT TO PREVENT PERMANENT ORGAN DAMAGE (ELEVATED BLOOD GLUCOSE LEVEL, PRESENCE OF KETONES AND SYMPTOMS E.G., NAUSEA, VOMITING, ABDOMINAL PAIN, CONFUSION), MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION DESCRIBE AND LOT 6001937 AND NO OTHER COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONLY ONE COMPLAINT RECEIVED ON THE LOT IN QUESTION AND HARM CODE, THE TEST ON REFERENCE SAMPLES WERE NOT TESTED SINCE A NO MALFUNCTION RELATED TO THE INFUSION SET WAS REPORTED, THEREFORE NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
E1: PATIENT CITY:(B)(6). PATIENT COUNTRY: CANADA.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT WENT TO THE HOSPITAL ON (B)(6) 2025 DUE TO HYPERGLYCEMIA EVENT. BLOOD GLUCOSE LEVEL WAS 32 MMOL/L AT THE TIME OF THE EVENT AND PATIENT GOT TREATED WITH INTRAVENOUS (IV) DRIP. PATIENT ALSO EXPERIENCED THE SYMPTOMS OF FEELING SICK AT THE TIME OF THE EVENT. PATIENT WAS FOUND POSITIVE FOR KETONES LEVEL. THE DURATION OF HOSPITALIZATION WAS 5 DAYS. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 220742 | QUICK-SET | UNO QUICK-SET 46/6 SC1 MCAN | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | MMT-394A600 | 6001937 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Hospitalization| R |