FDA Adverse Event Malfunction Summary report: N

INSET II

MDR report key: 22746671 · Received August 8, 2025

Report

Report Number
3003442380-2025-12615
Event Type
Malfunction
Date Received
August 8, 2025
Date of Event
July 19, 2025
Report Date
August 20, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6010640 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3.0 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2.0 FOR THE CODE OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). COMPLAINT INVESTIGATIONS: PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION XX ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL LEAK TEST 2 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, SAMPLE 6 AND SAMPLE 10 WERE FOUND LEAKING BY SUPERIOR MEMBRANE SAMPLES OUT 10 SAMPLES PASSED THE TEST. THIS MALFUNCTION ITS NOT RELEATED WITH THIS COMPLAINT. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010640 WAS MANUFACTURED ACCORDING TO THE WI 4905082 VERSION 75 IN THE INSET 6, ON 05/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 10/AUG/2025 AGAINST MALFUNCTION CODE OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED) AND LOT 6010640 AND NO OTHER COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT 6010640 AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO HARM, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE. NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN.THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN GERMANY. IT WAS REPORTED THAT THE PATIENT FACED AN INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS IN THE TUBING.THE INSERTION SITE WAS ABDOMEN. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
220740 INSET II UNO INSET II 46/6 GREY Y-CAP 10PAK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 86-046-52B6 6010640

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown