ALERE AFINION AS100 ANALYZER
Report
- Report Number
- 9613069-2025-00004
- Event Type
- Malfunction
- Date Received
- August 6, 2025
- Date of Event
- July 28, 2025
- Report Date
- November 12, 2025
- Manufacturer
- ABBOTT DIAGNOSTICS TECHNOLOGIES AS
- Product Code
- JQT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
EVENT IS RELATED TO SAME / SIMILAR DEVICE IN THE US UNDER PRODUCT NUMBER: 1115175 AND PRIMARY UDI: (B)(4).
EVENT IS RELATED TO SAME / SIMILAR DEVICE IN THE US UNDER PRODUCT NUMBER: 1115175 AND PRIMARY UDI: (B)(4). THE COMPLAINT DETAILS WERE REVIEWED ALONG WITH OUR INTERNAL RECORDS AND NO NONCONFORMANCES WERE IDENTIFIED. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE REPORTED IN THE COMPLAINT. THERE IS NO INDICATION THAT THE REPORTED ISSUES IS RELATED TO PRODUCT RESULTS, INADEQUATE INSTRUCTIONS FOR USE OR INACCURATE LABELING. FINDINGS FROM INTERNAL INVESTIGATION COULD NOT CONFIRM ISSUES RELATED TO SPARK ON THE ANALYZER ITSELF. HOWEVER, THE SELF-TEST ERROR COULD BE REPRODUCED INTERNALLY, AND IT APPEARS TO BE LINKED TO A SLOW FLIP-LID MECHANISM. THE SELF-TEST VALIDATES HARDWARE AND SOFTWARE INTEGRITY, TEST CARTRIDGE TRANSPORT SYSTEM, LIQUID TRANSPORT SYSTEM, AND THE CAMERA VISION SYSTEM. WHEN THE ANALYZER IS POWERED ON FOR A LONGER PERIOD, IT WILL AUTOMATICALLY RESTART ONCE A DAY TO ENSURE THAT THE SELF-TEST IS DONE REGULARLY. USERS MAY RECEIVE A MESSAGE, ¿SELF-TEST ERROR; ANALYZER IN NON-OPERATIVE MODE¿. FAILURE TO COMPLETE THE SELF-TEST CAN BE CAUSED BY A VARIETY OF POTENTIAL ROOT CAUSES AND IF THE ISSUE PERSISTS THE ANALYZER NEEDS TO BE RETURNED AND REPLACED. CUSTOMER COMPLAINTS WILL CONTINUE TO BE MONITORED TO ENSURE CONTROL LIMITS ARE NOT EXCEEDED AND ENSURE THAT PRODUCT PERFORMANCE MEETS CUSTOMER EXPECTATIONS.
ON 28-JUL-2025 A CUSTOMER REPORTED TO HAVE AN ISSUE WITH THEIR AFINION AS100 ANALYZER AFTER THEY PERFORMED THE CLEANING. TECHNICAL SERVICES CONTACTED THE CUSTOMER BY PHONE, WHO REPORTED THAT AFTER THEY CONNECTED THEIR AFINION AS100 ANALYZER TO THE POWER OUTLET AFTER CLEANING THE INSTRUMENT, IT DISPLAYED A SELF-TEST ERROR. THE CUSTOMER THEN PULLED OUT THE POWER SUPPLY AGAIN AND SHORTLY BEFORE PLUGGING IN THE POWER CABLE INTO THE WALL AGAIN IT SPARKED. ON THE SAME DATE THE CUSTOMER ALSO CONFIRMED: - ONLY THE USER OF THE INSTRUMENT WAS INVOLVED AND NOT IMPACTED. - THE SPARK CAME FROM THE POWER SUPPLY AND THE CIRCUIT BREAKER DID NOT TRIP. - THE USER DID NOT ATTEMPT TO USE THE INSTRUMENT AFTER THE ISSUE OCCURRED. - THE ANALYZER IS PLACED WITH SUFFICIENT SURROUNDING AIRSPACE (AT LEAST 10 CM ON EACH SIDE), AS PER THE USER MANUAL. - NO OTHER OBSERVATIONS WERE MADE BY THE CUSTOMER.
ON 28-JUL-2025 A CUSTOMER REPORTED TO HAVE AN ISSUE WITH THEIR AFINION AS100 ANALYZER AFTER THEY PERFORMED THE CLEANING. TECHNICAL SERVICES CONTACTED THE CUSTOMER BY PHONE, WHO REPORTED THAT AFTER THEY CONNECTED THEIR AFINION AS100 ANALYZER TO THE POWER OUTLET AFTER CLEANING THE INSTRUMENT, IT DISPLAYED A SELF-TEST ERROR. THE CUSTOMER THEN PULLED OUT THE POWER SUPPLY AGAIN AND SHORTLY BEFORE PLUGGING IN THE POWER CABLE INTO THE WALL AGAIN IT SPARKED. ON THE SAME DATE THE CUSTOMER ALSO CONFIRMED: - ONLY THE USER OF THE INSTRUMENT WAS INVOLVED AND NOT IMPACTED. - THE SPARK CAME FROM THE POWER SUPPLY AND THE CIRCUIT BREAKER DID NOT TRIP. - THE USER DID NOT ATTEMPT TO USE THE INSTRUMENT AFTER THE ISSUE OCCURRED. - THE ANALYZER IS PLACED WITH SUFFICIENT SURROUNDING AIRSPACE (AT LEAST 10 CM ON EACH SIDE), AS PER THE USER MANUAL. - NO OTHER OBSERVATIONS WERE MADE BY THE CUSTOMER. ON 31-JUL-2025 THE CUSTOMER FURTHER CONFIRMED: - THE SPARK WAS FROM THE POWER SUPPLY. - IT WAS FROM THE MAINS CORD. - THERE WAS NO SHOCK OR ELECTROSTATIC DISCHARGE, ONLY A SPARK. - THE CIRCUIT BREAKER DID NOT TRIP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1365548 | ALERE AFINION AS100 ANALYZER | ANALYZER | JQT | ABBOTT DIAGNOSTICS TECHNOLOGIES AS | 10187763 | 10187763 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |