FDA Adverse Event Malfunction Summary report: N

OPTICROSS 18

MDR report key: 22725325 · Received August 6, 2025

Report

Report Number
2124215-2025-54055
Event Type
Malfunction
Date Received
August 6, 2025
Date of Event
July 14, 2025
Report Date
September 23, 2025
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
OBJ
PMA / PMN Number
K160514
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATED BY MANUFACTURER: THE DEVICE WAS RETURNED FOR ANALYSIS. VISUAL INSPECTION SHOWS THE SHEATH AND IMAGING WINDOW WERE KINKED. MICROSCOPIC INSPECTION REVEALED THE GUIDEWIRE EXIT PORT WAS LIFTED, BUT THE DISTAL SECTION OF THE TIP WAS IN GOOD CONDITION. A TEST GUIDEWIRE WAS INSERTED, AND NO INDICATION OF RESISTANCE IN TRACKING THE GUIDEWIRE INTO THE CATHETER WAS NOTED. G4: PREMARKET / 510(K) #: K160514, K222568.

Additional Manufacturer Narrative · 0

G4: PREMARKET / 510(K) #: K160514, K222568. G1 MFR CONTACT FIRST NAME, MFR CONTACT LAST NAME: CORRECTION.

Additional Manufacturer Narrative · 0

G4: PREMARKET / 510(K) #: K160514, K222568.

Description of Event or Problem · 0

IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED PERIPHERAL ARTERY. AN OPTICROSS 18 IMAGING CATHETER WAS SELECTED FOR USE IN THE ULTRASOUND EXAMINATION. DURING THE PROCEDURE, THE CATHETER BECAME KINKED MID SHAFT AND WRAPPED AROUND THE NON-BOSTON SCIENTIFIC GUIDEWIRE. THE CATHETER WAS EXCHANGED, AND THE SAME THING HAPPENED WITH A NEW OPTICROSS 18 IMAGING CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED PERIPHERAL ARTERY. AN OPTICROSS 18 IMAGING CATHETER WAS SELECTED FOR USE IN THE ULTRASOUND EXAMINATION. DURING THE PROCEDURE, THE CATHETER BECAME KINKED MID SHAFT AND WRAPPED AROUND THE NON-BOSTON SCIENTIFIC GUIDEWIRE. THE CATHETER WAS EXCHANGED, AND THE SAME THING HAPPENED WITH A NEW OPTICROSS 18 IMAGING CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED PERIPHERAL ARTERY. AN OPTICROSS 18 IMAGING CATHETER WAS SELECTED FOR USE IN THE ULTRASOUND EXAMINATION. DURING THE PROCEDURE, THE CATHETER BECAME KINKED MID SHAFT AND WRAPPED AROUND THE NON-BOSTON SCIENTIFIC GUIDEWIRE. THE CATHETER WAS EXCHANGED, AND THE SAME THING HAPPENED WITH A NEW OPTICROSS 18 IMAGING CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1765753 OPTICROSS 18 CATHETER, ULTRASOUND, INTRAVASCULAR OBJ BOSTON SCIENTIFIC CORPORATION H7493932800180 0036162469

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ABBOTT SPARTACORE GUIDEWIRE| ABBOTT SPARTACORE GUIDEWIRE| ABBOTT SPARTACORE GUIDEWIRE