FDA Adverse Event Injury Summary report: N

QUICK-SET

MDR report key: 22724953 · Received August 6, 2025

Report

Report Number
3003442380-2025-12449
Event Type
Injury
Date Received
August 6, 2025
Date of Event
July 11, 2025
Report Date
July 23, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244017573
PMA / PMN Number
K991759
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED FOR RECORD (B)(4) ON 23-JUL-2025. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011504 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 ON 03-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED TRENDING: A QUERY WAS RUN IN DATABASE ON 23-JUL-2025 AGAINST HARM CODE SIGNS OF UNTREATED HYPOGLYCEMIA THAT PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY A HEALTH CARE PROFESSIONAL (HCP) OR REQUIRES EMERGENCY ADVANCED LIFE, MALFUNCTION CODE NO MALFUNCTION DESCRIBE AND LOT 6011504 AND NO MORE COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONLY ONE COMPLAINT RECEIVED ON THE LOT IN QUESTION AND HARM CODE, THE TEST ON REFERENCE SAMPLES WERE NOT TESTED SINCE A NO MALFUNCTION RELATED TO THE INFUSION SET WAS REPORTED, THEREFORE NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

E1: PATIENT CITY - (B)(6). PATIENT COUNTRY - UNITED STATES. THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED FOR RECORD COMPLAINT (B)(4) ON (B)(6) 2025. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. SEE ATTACHMENT COMPLAINT (B)(4) COMPLAINT TEST REPORT.PDF FOR THE DETAILS. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011504 WAS MANUFACTURED ACCORDING TO THE[?]WORK INSTRUCTION (WI) VERSION 82 ON (B)(6) 2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED TRENDING: A QUERY WAS RUN IN DATABASE ON (B)(6) 2025 AGAINST HARM CODE SIGNS OF UNTREATED HYPOGLYCEMIA THAT PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY A HEALTH CARE PROVIDER (HCP) OR REQUIRES EMERGENCY ADVANCED LIFE, MALFUNCTION CODE NO MALFUNCTION DESCRIBE AND LOT 6011504 AND NO MORE COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONLY ONE COMPLAINT RECEIVED ON THE LOT IN QUESTION AND HARM CODE, THE TEST ON REFERENCE SAMPLES WERE NOT TESTED SINCE A NO MALFUNCTION RELATED TO THE INFUSION SET WAS REPORTED, THEREFORE NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED LOW BLOOD GLUCOSE LEVELS AND LATER GOT AMBULATORY ASSISTANCE FOR THE TREATMENT ON (B)(6) 2025. THE BLOOD GLUCOSE LEVEL WAS 48 MG/DL AT THE TIME OF EVENT. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1337203 QUICK-SET UNO QUICK-SET 60/9 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-397A 6011504 05705244017573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown