INSET II
Report
- Report Number
- 3003442380-2025-12420
- Event Type
- Malfunction
- Date Received
- August 6, 2025
- Date of Event
- May 1, 2025
- Report Date
- September 27, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018129
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR (B)(4). DEVICE 2 OF 3.
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2025-12420. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: REVISION 21 OF 3709030 DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF 3709030. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010193, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 26-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6010193". THE COUNT OF COMPLAINT IS 0 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010193 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN THE LINE 49 ON 16/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. GLUE-TUBING LOT: THE LOT 4L00272 WAS MANUFACTURED ACCORDING TO THE WI VERSION 8 AND MANUFACTURED IN THE MACHINE ITL01 ON 11/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT 4L00282 WAS MANUFACTURED ACCORDING TO THE WI VERSION 8 AND MANUFACTURED IN THE MACHINE ITL01 ON 14/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT 4L02714 WAS MANUFACTURED ACCORDING TO THE WI VERSION 65 AND MANUFACTURED IN THE MACHINE SC03 ON 13/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT A NON-CONFORMANCE NC WAS OPENED DURING THE STERILIZATION PROCESSES ACTIONS WERE REQUIRED. THEREFORE, DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED EVENTS WHERE THREE INFUSION SETS TUBING WAS BROKEN IN HALF BETWEEN (B)(6) 2025 .THE INFUSION SETS WERE IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1364728 | INSET II | UNO INSET II 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002817 | 6010193 | 05705244018129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR | Male |