FDA Adverse Event Injury Summary report: N

SMS STEM

MDR report key: 22714426 · Received August 5, 2025

Report

Report Number
3005180920-2025-00730
Event Type
Injury
Date Received
August 5, 2025
Date of Event
July 21, 2025
Report Date
August 5, 2025
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030888854
PMA / PMN Number
K181693
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 22-07-2025. STEM: SMS SOLID 01.36.073 SMS SOLID STEM LAT SIZE 13 (K181693) LOT 1903494: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05-07-2019 EXPIRATION DATE: 2024-06-24. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION PERFORMED BY CLINICAL AFFAIRS DEPARTMENT: REVISION SURGERY WAS PERFORMED APPROXIMATELY ONE YEAR AFTER THE PRIMARY TOTAL HIP ARTHROPLASTY (THA) DUE TO STEM LOOSENING. THE PATIENT HAD REPORTED PAIN AND UNDERWENT RADIOLOGICAL FOLLOW-UP. THE AVAILABLE X-RAY IMAGES CONFIRM STEM MOBILIZATION, WITH RADIOLUCENT LINES PARTICULARLY EVIDENT AROUND THE SHOULDER OF THE STEM. ASEPTIC LOOSENING IS A WELL-DOCUMENTED COMPLICATION IN THE LITERATURE, OFTEN WITH MULTIFACTORIAL OR UNCLEAR ETIOLOGY. EVEN IN THIS CASE, THE PRECISE CAUSE OF THE FAILURE REMAINS DIFFICULT TO DETERMINE BASED ON THE AVAILABLE INFORMATION. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE-DESCRIBED ADVERSE EVENT AFTER PRIMARY HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. NO PREVIOUS CASE HAS BEEN ADDRESSED TO A DEVICE DESIGN OR MANUFACTURING RELATED ROOT CAUSE.

Description of Event or Problem · 0

AT ABOUT 1 YEAR AND 2 MONTHS FROM PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE STEM, HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2684762 SMS STEM SMS SOLID STEM LAT SIZE 13 LZO MEDACTA INTERNATIONAL SA 01.36.073 1903494 07630030888854

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male Required Intervention