FDA Adverse Event Malfunction Summary report: N

VERASEAL DUAL APPL 35CM R CE

MDR report key: 22698596 · Received August 4, 2025

Report

Report Number
2210968-2025-08960
Event Type
Malfunction
Date Received
August 4, 2025
Date of Event
January 1, 2025
Report Date
March 20, 2026
Manufacturer
ETHICON INC.
Product Code
MZM
UDI-DI
10705031465473
PMA / PMN Number
BK190324
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). H6. COMPONENT CODE: G07002 - DEVICE NOT RETURNED. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION WAS REQUESTED, AND THE FOLLOWING WAS OBTAINED: - WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO. - WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. - WERE FRAGMENTS GENERATED? NO. - IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN. - PATIENT STATUS/ OUTCOME / CONSEQUENCES NO. THE FOLLOWING INFORMATION WAS REQUESTED, BUT UNAVAILABLE: - WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN. - IS THE PATIENT PART OF A CLINICAL STUDY UNKNOWN. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. 1. IS THE USER A NEW USER TO VERASEAL? IF NOT, HOW MANY TIMES HAVE THEY USED VERASEAL PRIOR TO THIS EVENT? 2. HOW MANY TIMES HAVE THEY APPLIED THE LAPAROSCOPIC TIP? 3. WAS A SALES REPRESENTATIVE PRESENT DURING THE CASE WHEN THE ISSUE WAS EXPERIENCED? 4. WAS ANY LEAKAGE OR PRODUCT SOLUTION OBSERVED AT THE LUER LOCKS CONNECTION? 5. WERE THERE ANY UNEXPECTED OUTCOMES OR COMPLICATIONS AS A RESULT OF THE EVENT? 6. ARE THERE PICTURES OF THE DAMAGED DEVICE AVAILABLE? THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT #(B)(4), CORRECTED INFORMATION: H 3. DEVICE EVALUATED BY MANUFACTURER? ADDITIONAL INFORMATION: H 6. COMPONENT CODE, H 6. TYPE OF INVESTIGATION, H 6. INVESTIGATION FINDINGS, H 6. INVESTIGATION CONCLUSIONS H3 INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE VRAL35 DEVICE WAS RETURNED WITH CANNULA SHAFT BENT AND ONE LUER LOCK ATTACHED. LUER LOCK DAMAGED. IN ADDITION, THE OPENED PACKAGING WAS RETURNED. DUE TO THE CONDITION OF THE LUER LOCK ATTACHED, NO FUNCTIONAL TESTING COULD BE PERFORMED TO EVALUATE THE REPORTED INCIDENT. AS PART OF ETHICON¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE EVENT REPORTED WAS CONFIRMED AND IT IS RELATED TO IMPROPER USE OF THE DEVICE. ONE POSSIBLE CAUSE FOR THE DAMAGE FOUND ON THE ADAPTER MAY BE EXCESSIVE EXTERNAL LOAD PLACED ON THE DEVICE. A REVIEW OF THE DEVICE BATCH RECORD WAS PERFORMED FOR RAW MATERIALS, IN-PROCESS, STERILIZATION, AND FINISHED DEVICES FOR LOT NUMBER 5513930 AND THERE WERE NO NON-CONFORMANCES OR CAPAS OPENED IN RELATION TO THE NATURE OF THE ALLEGED COMPLAINT. THE DEVICES MET ALL RAW MATERIALS, IN-PROCESS, STERILIZATION AND FINISHED DEVICE SPECIFICATIONS UPON RELEASE OF THE PRODUCT. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT #(B)(4). CORRECTED INFORMATION: D 9. DEVICE AVAILABLE FOR EVALUATION? ADDITIONAL INFORMATION: D 9. DATE DEVICE RETURNED TO MANUFACTURER, D 9. IS DEVICE RETURNED TO MANUFACTURER?, H 6. COMPONENT CODE, H 6. TYPE OF INVESTIGATION, H 6. INVESTIGATION FINDINGS, H 6. INVESTIGATION CONCLUSIONS. H6 COMPONENT CODE: G07002 - PENDING EVALUATION OF RETURNED DEVICE. THE DEVICE UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A FIBRIN SEALANT PREPARATION DEVICE WAS USED. THE SCRUB NURSE WAS HAVING AN ISSUE ATTACHING THE LAPAROSCOPIC APPLICATOR TO THE FIBRIN SEALANT PREPARATION DEVICE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204959 VERASEAL DUAL APPL 35CM R CE FIBRIN SEALANT PREPARATION DEVICE MZM ETHICON INC. 5513930 10705031465473

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown