FDA Adverse Event Malfunction Summary report: N

BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD

MDR report key: 22698185 · Received August 4, 2025

Report

Report Number
1119779-2025-02325
Event Type
Malfunction
Date Received
August 4, 2025
Date of Event
December 22, 2023
Report Date
June 19, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JSJ
UDI-DI
10382902212772
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: DURING MANUFACTURING OF MATERIAL 221277, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD REVIEW FOR BATCH 3296800 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. ALL BATCHES ARE TESTED PRIOR TO RELEASE AND RESULTS REPORTED ON THE CERTIFICATE OF ANALYSIS WHICH CAN BE OBTAINED AT WWW.BD.COM/REGDOCS. PHENYLETHYL ALCOHOL W/ 5% SB IS STABILITY TESTED BIENNIALLY FOR BIOLOGICAL PERFORMANCE TO ENSURE SATISFACTORY PERFORMANCE THROUGHOUT SHELF LIFE WITH THE ORGANISMS THAT ARE REPORTED ON THE CERTIFICATE OF ANALYSIS. ALL PERFORMANCE TESTING ON THIS BATCH WAS SATISFACTORY AT THE TIME OF RELEASE. THE COMPLAINT HISTORY WAS REVIEWED; THERE ARE NO OTHER COMPLAINTS TAKEN ON THIS BATCH 3296800 FOR THIS DEFECT. THERE WERE NO RETENTION SAMPLES AVAILABLE FOR INVESTIGATION OF THIS COMPLAINT. TWO PHOTOS WERE RECEIVED, ONE PHOTO SHOWS THE PLATE PRINT: 3296800, TIME STAMP 23:46. THE SECOND PHOTO SHOWS A VISIBLY STREAKED PLATE WITH GROWTH. OBSERVATIONS ABOUT PERFORMANCE CANNOT BE MADE FROM PHOTOS. NO RETURNS OR RETAINS WERE AVAILABLE FOR INVESTIGATION OF THIS COMPLAINT. THIS COMPLAINT CANNOT BE CONFIRMED. NO COMPLAINT TRENDS FOR THIS DEFECT HAVE BEEN IDENTIFIED; NO ACTIONS ARE INDICATED AT THIS TIME. BD WILL CONTINUE TO TREND COMPLAINTS FOR PERFORMANCE. THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING APRIL 2023-2025, UNDER CAPA 11910483.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202253 BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL JSJ BECTON DICKINSON & CO. (SPARKS) 3296800 10382902212772

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown