FDA Adverse Event Malfunction Summary report: N

QUICK-SET

MDR report key: 22696954 · Received August 4, 2025

Report

Report Number
3003442380-2025-12242
Event Type
Malfunction
Date Received
August 4, 2025
Date of Event
July 12, 2025
Report Date
November 7, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244017573
PMA / PMN Number
K991759
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DEVICE HISTORY RECORD (DHR) REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010412, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010412 WAS MANUFACTURED ACCORDING TO TH E WORK INSTRUCTION (WI) VERSION 82 AND PACKAGING IN THE MACHINE MULTIVAC 12 ON 25-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 4L01889 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 25-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 4L01890 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 25-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 4L05063 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 25-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 4L03187 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE MP04-MP08, ON 21-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 4L03189 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE 04-08, ON 23-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE; NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. EXTENDED INSPECTION IS NOT RELATED WITH THE CLAIMED MALFUNCTION THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE EVENT ON (B)(6) 2025. THE LEAKAGE WAS AT THE QUICK RELEASE (QR). THE INFUSION SET WAS IN USE FOR ONE DAY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
390767 QUICK-SET UNO QUICK-SET 60/9 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-397A 6010412 05705244017573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown