FDA Adverse Event Injury Summary report: N

GLIDEWELL HT IMPLANT Ø4.3 X 8 MM

MDR report key: 22691486 · Received August 1, 2025

Report

Report Number
3011649314-2025-00870
Event Type
Injury
Date Received
August 1, 2025
Date of Event
July 14, 2025
Report Date
December 29, 2025
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT DEVICE HAS BEEN RETURNED AND THE INVESTIGATION IS CURRENTLY ONGOING. ONCE THE RETURNED DEVICE HAS BEEN EVALUATED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED. THE INVESTIGATION HAS BEEN UPDATED, AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE STOCK PRODUCT FOR GLIDEWELL HT IMPLANT LOT# 6232218 WAS REVIEWED AND FOUND TO BE IN STOCK, BUT IT IS NOT APPLICABLE FOR PHYSICAL EVALUATION SINCE THE ISSUE IS CUSTOMER RELATED. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN THE ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A GLIDEWELL HT IMPLANT Ø4.3 X 8 MM (70-1189-IMP0009) USING THE RADIOGRAPHIC TEMPLATE (MKT-013579 REV 1 PK-4500230-112023). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE TREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE DESCRIPTION: "LACK OF PRIMARY STABILITY" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, THE PATIENT HAS TYPE II BONE QUALITY. IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "GLIDEWELL HT IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE" IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." PER THE REPORTED INFORMATION, THERE WAS A FIT ISSUE AS WELL. THE PROBABLE CAUSE FOR THIS ISSUE MAY BE DUE TO AN OVER PREPARED OSTEOTOMY. IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE SITE PREPARATION SECTION: "IF PLACING A GLIDEWELL HT IMPLANT THAT IS 3.5 MM IN DIAMETER OR GREATER, SHAPING DRILLS ARE USED SEQUENTIALLY TO WIDEN THE OSTEOTOMY TO THE MATCHING DIAMETER. TO AVOID OVER-PREPARATION, WIDENING DRILL DIAMETERS SHOULD BE USED ONLY AS NEEDED, AND IN PROPER SUCCESSION. EACH SHAPING DRILL IS LENGTH-SPECIFIC, TO MATCH THE LENGTH OF THE PRESCRIBED IMPLANT. OSTEOTOMY DEPTH MAY BE INCREASED SEQUENTIALLY, BEGINNING WITH SHORTER DRILL LENGTHS, PROVIDED SUFFICIENT DEPTH IS ACHIEVED WITH THE FINAL DRILL. SELECT THE DESIRED SHAPING DRILL, ACCOUNTING FOR BONE DENSITY AND THE SIZE OF THE IMPLANT TO BE PLACED. WITH COPIOUS IRRIGATION, DRILL TO DEPTH. THE FINAL DRILL SHOULD CORRESPOND WITH THE MATCHING IMPLANT SIZE, AS CHARTED BELOW, WITH THE GOAL OF ACHIEVING HIGH PRIMARY STABILITY UPON IMPLANT PLACEMENT." THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT A GLIDEWELL HT IMPLANT FAILED. THE PATIENT'S BONE QUALITY IS TYPE II AND THEIR ORAL HYGIENE IS GOOD. THE PATIENT PRESENTED ON (B)(6) 2025 FOR A PRIMARY PROCEDURE ON TOOTH #18. ON (B)(6) 2025, AFTER HEALING ABUTMENT PLACEMENT THE PROVIDER DETERMINED THAT THE IMPLANT LACKED PRIMARY STABILITY AND REMOVED THE IMPLANT. NO PERMANENT INJURY WAS REPORTED. PER THE REPORTED INFORMATION, THE PATIENT HAS NO PREEXISTING MEDICAL CONDITIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1115173 GLIDEWELL HT IMPLANT Ø4.3 X 8 MM GLIDEWELL HT IMPLANT SYSTEM DZE PRISMATIK DENTALCRAFT, INC. 70-1189-IMP0009 6232218

Patients

Seq Age Sex Outcome Treatment
1 64 YR Male Required Intervention