FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 2268610 · Received September 28, 2011

Report

Report Number
9611451-2011-00607
Event Type
Malfunction
Date Received
September 28, 2011
Report Date
August 30, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. LOT NUMBERS: 110414: X1, 110620: X1. DEVICE MANUFACTURING DATES: 04/14/2011, 06/20/2011. METHOD: TWO COMPLAINT DEVICES HAVE BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR VISUAL INSPECTION AND PERFORMANCE TEST. RESULT: THE VISUAL INSPECTION REVEALED NO DAMAGES FOUND ON BOTH COMPLAINT DEVICES. THE PERFORMANCE TEST REVEALED THAT THERE WAS CONDENSATION IN THE EVAQUA2 LIMBS WHEN THE HEATER WIRE WAS NOT CONNECTED, BUT NO CONDENSATION WAS OBSERVED WHEN THE HEATER WIRE WAS CONNECTED. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS TYPE FOR LOT NUMBERS 110414 AND 110620. CONCLUSION: CONDENSATE IN THE HUMIDIFICATION SYSTEM, ALTHOUGH NOT PREFERRED, IS AN EXPECTED SIDE EFFECT OF HEATED PASS-OVER HUMIDIFICATION SYSTEMS IN MANY CONDITIONS, AND MAY VARY BETWEEN LIGHT MISTING TO WATER DROPLETS THAT FORM ON THE WALL OF COOL BREATHING CIRCUIT TUBING. THE AMOUNT OF CONDENSATE IN THE VENTILATION SYSTEM IS INFLUENCED BY A NUMBER OF MULTIPLE SETUP AND ENVIRONMENTAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. ONE COMPLAINT DEVICE HAS RECENTLY BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND THE DEVICE EVALUATION IS CURRENTLY IN PROGRESS. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE SALES REPRESENTATIVE THAT EXCESSIVE CONDENSATION WAS FOUND IN THE EXPIRATORY LIMB OF TWO RT265 INFANT BREATHING CIRCUIT KITS. AN INCREASE IN THE CO2 LEVELS OF THE PATIENTS WAS OBSERVED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE SALES REPRESENTATIVE THAT EXCESSIVE CONDENSATION WAS FOUND IN THE EXPIRATORY LIMB OF TWO RT265 INFANT BREATHING CIRCUIT KITS. AN INCREASE IN THE CO2 LEVELS OF THE PATIENTS WAS OBSERVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT265 SEE H10

Patients

Seq Age Sex Outcome Treatment
1