FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 SYNCHRON® SYSTEM

MDR report key: 2268290 · Received September 28, 2011

Report

Report Number
2050012-2011-05630
Event Type
Malfunction
Date Received
September 28, 2011
Date of Event
September 3, 2011
Report Date
September 4, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(6) 2011 AND FOUND FOREIGN DEBRIS CLOGGING THE ISE DRAIN VALVE. FSE CLEANED THE VALVE AND VERIFIED PERFORMANCE ACCORDING TO ESTABLISHED PROCEDURES. FSE ALSO ENCOUNTERED INTERMITTENT REAGENT LEVEL SENSE ERROR WHILE TROUBLESHOOTING AND PROCEEDED TO REPLACE AND ALLIGN PROBE A LEVEL SENSE BEAD. INSTRUMENT REPAIRS WERE AGAIN VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

CUSTOMER CALLED REPORTING OF AN ERRONEOUS ION SELECTIVE ELECTRODES (ISE) RESULT GENERATED FROM A UNICEL DXC 600 SYNCHRON SYSTEM ON (B)(6) 2011. CUSTOMER INDICATED THAT THE ISE MODULE IS A POST ISE MODIFICATION MODULE. CUSTOMER REPORTED THAT THE INSTRUMENT GENERATED A FALSE HIGH NA , A FALSE LOW CL AND A SUPPRESSED CALC RESULTS FOR ONE PATIENT SAMPLE WHILE K AND CO2 RESULTS WERE NOT AFFECTED. CUSTOMER STATED THAT THE PATIENT'S SAMPLE HAS BEEN RUN ON ANOTHER INSTRUMENT (UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM) AND THE ORIGINAL RESULTS WERE AMENDED. INSTRUMENT PRINTOUTS INDICATED THAT THE ORIGINAL INSTRUMENT IN QUESTION GENERATED A RESULT OF 154 MMOL/L OF NA, 76 MMOL/L OF CL AND A SUPPRESSED CALC RESULT DUE TO A CALIBRATION REFERENCE DRIFT. THE REPEAT ON ANOTHER INSTRUMENT (DXC 600I) GAVE A RESULT OF 139 MMOL/L OF NA, 104 MMOL/L OF CL AND 8.6 MG/DL OF CALC. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LAB.