FDA Adverse Event Injury Summary report: N

XPS STRAIGHTSHOT M4 MICRODEBRIDER

MDR report key: 2268084 · Received September 28, 2011

Report

Report Number
1045254-2011-00073
Event Type
Injury
Date Received
September 28, 2011
Date of Event
August 30, 2011
Report Date
December 21, 2011
Manufacturer
MEDTRONIC XOMED, INC.
Product Code
ERL
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). PMA/510(K) IS K041413. NO MEDWATCH 3500A FORM WAS RECEIVED FROM THE REPORTER. ANY MISSING OR INCOMPLETE DATA ON THIS FORM 3500A IS THE RESULT OF INFORMATION NOT BEING PROVIDED OR RELEASED BY THE REPORTER. MULTIPLE ATTEMPTS TO OBTAIN THE REQUIRED INFORMATION WERE MADE, AND THE RECORDS OF THESE ATTEMPTS ARE DOCUMENTED IN THE COMPLAINT FILE. THIS PRODUCT WAS BEING USED FOR TREATMENT, NOT DIAGNOSIS.

Additional Manufacturer Narrative · 1

(B)(4). THE USER GUIDE WARNS THAT USE OF A BUR WITHOUT IRRIGATION MAY CAUSE AN INORDINATE AMOUNT OF HEAT BUILDUP RESULTING IN THERMAL INJURY TO TISSUE. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT REMOVAL OF AN OSTEOMA FROM THE FRONTAL SINUS VIA A LOTHROP PROCEDURE. THE SURGEON USED A MICRODEBRIDER WITH THE 55 DEGREE DRILL ATTACHMENT AND HELD THE DRILL SHAFT RATHER THAN THE HANDLE OF THE MICRODEBRIDER. THE DEVICE WAS USED AT 12000 RPMS IN FORWARD DIRECTION. THINGS WERE GOING WELL UNTIL THE PHYSICIAN FELT THE DRILL HEAT UP. THE PHYSICIAN STOPPED IMMEDIATELY. IT WAS REPORTED THAT THE FLUID BAG AND IRRIGATION HAD RUN OUT. THE PATIENT RECEIVED A 1ST DEGREE BURN TO HER UPPER LIP, JUST BELOW THE NOSTRIL. THE PATIENT WAS TREATED WITH ANTIBIOTIC CREAM FOR THE BURN (SILVADENE). THE PATIENT IS CURRENTLY DOING FINE. REPORTEDLY, THE PATIENT RECEIVED A SUPERFICIAL BURN THAT SHOULD HOPEFULLY HEAL WITHOUT A SCAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XPS STRAIGHTSHOT M4 MICRODEBRIDER DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL MEDTRONIC XOMED, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 00048 YR Required Intervention (B)(4)