ENDOWRIST
Report
- Report Number
- 2955842-2025-32345
- Event Type
- Malfunction
- Date Received
- July 30, 2025
- Date of Event
- July 10, 2025
- Report Date
- July 14, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874119808
- PMA / PMN Number
- K214095
- Removal / Correction Number
- ISIFA2024-09-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CI
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN GRIP CABLE. WHEN CABLE BREAKAGE OCCURS, THE INSTRUMENT IS IMMEDIATELY INOPERABLE DUE TO LOSS OF FUNCTIONALITY. THE INSTRUMENT HAS TUNGSTEN DRIVE CABLES TO TRANSMIT MOTION FROM THE INPUT DISCS IN THE HOUSING, THROUGH THE MAIN SHAFT, AND TO THE DISTAL INSTRUMENT TIP. THESE CABLES ARE EXPOSED AT THE INSTRUMENT TIP AND CONTROL MOVEMENTS AT THE WRIST. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE. THE PROBABLE ROOT CAUSE OF CABLE BREAKAGE, WHETHER PARTIAL OR FULL, IS ATTRIBUTED TO DAMAGE TO THE CABLE THROUGH EXTERNAL COLLISIONS, DURING USE, OR DURING REPROCESSING. THIS ISSUE CAN BE RESOLVED BY USING AN ALTERNATE INSTRUMENT TO COMPLETE THE PROCEDURE.
AN INVESTIGATION IS IN PROGRESS TO DETERMINE THE CAUSE AN RMA WAS ISSUED TO EVALUATE THE INTUITIVE SURGICAL, INC. (ISI) DEVICE. OF THIS REPORTED EVENT. ADDITIONAL INFORMATION IS BEING GATHERED TO DETERMINE THE CONTRIBUTION OF THE DEVICE TO THE CUSTOMER REPORTED ISSUE.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INSTRUMENT STOPPED WORKING PROPERLY, SO IT WAS REMOVED FOR EXAMINATION. IT WAS OBSERVED THAT ONE OF THE PULLEYS CABLES WAS CUT, SO IT WAS REPLACED WITH A BACKUP INSTRUMENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55175 | ENDOWRIST | FENESTRATED BIPOLAR FORCEPS | NAY | INTUITIVE SURGICAL, INC | 471205-17 | K10241227 0048 | 00886874119808 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |