FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 22652085 · Received July 29, 2025

Report

Report Number
2955842-2025-32199
Event Type
Malfunction
Date Received
July 29, 2025
Date of Event
June 29, 2025
Report Date
July 9, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874120767
PMA / PMN Number
K214095
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL INC. (ISI), HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE FORCE BIPOLAR WAS ANALYZED AND FOUND TO HAVE A BROKEN CONDUCTOR WIRE WITHIN THE BIPOLAR YAW PULLEY, BETWEEN THE WIRE CRIMP ON THE GRIP AND THE SILICONE POTTING, WHICH SEALS THE WIRE ENTRANCE TO THE YAW PULLEY. THE INSTRUMENT FAILED THE ELECTRICAL CONTINUITY TEST, CONFIRMING A COMPLETE BREAK IN THE WIRE. NO THERMAL DAMAGE WAS FOUND ON THE INSTRUMENT. COMPONENTS ADJACENT TO THE BROKEN DO NOT SHOW DAMAGE. THE COMPLAINT REGARDING ONE OF THE BLACK CORDS SNAPPED WAS CONFIRMED BY FAILURE ANALYSIS. COMMON CAUSES OF A BROKEN BIPOLAR CONDUCTOR WIRE ARE ATTRIBUTED TO DAMAGE THROUGH EXTERNAL COLLISIONS, DURING USE, OR DURING REPROCESSING.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING CENTRAL PROCESSING, ONE OF THE BLACK CABLES OF THE FORCE BIPOLAR INSTRUMENT SNAPPED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE DAMAGE WAS IDENTIFIED DURING DECONTAMINATION. THE BLACK CORDS FOR THE ARTICULATION OF THE JAW SNAPPED. THERE WAS NO LOSS OF CAUTERY DUE TO THE BROKEN CABLE. THERE WAS NO SIGN OF THERMAL DAMAGE. THE PROCEDURE WAS COMPLETED USING THE SAME INSTRUMENT. THERE IS O REPORT OF FRAGMENTS FALLING INTO THE PATIENT'S ANATOMY.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1356490 ENDOWRIST FORCE BIPOLAR NAY INTUITIVE SURGICAL, INC 471405-06 K10241017 0012 00886874120767

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female DA VINCI INSTRUMENTS AND ACCESSORIES