BD SYRINGE 10ML LL BNS
Report
- Report Number
- 1213809-2025-00500
- Event Type
- Malfunction
- Date Received
- July 29, 2025
- Date of Event
- June 20, 2025
- Report Date
- September 19, 2025
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- PMA / PMN Number
- FMF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
(B)(4) FOLLOW UP FOR DEVICE EVALUATION. ONE PHOTO AND ONE SAMPLE OF A 10 ML LUER-LOK SYRINGE (PART NUMBER 301029) WERE RECEIVED AND EVALUATED. THE SAMPLE, RECEIVED FULLY ASSEMBLED AND LOOSE, EXHIBITED AN UNKNOWN BROWN FOREIGN MATERIAL INSIDE THE FLUID PATH. THE PHOTO PROVIDED WAS CONSISTENT WITH THE CONDITION OBSERVED DURING SAMPLE EVALUATION. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE FOREIGN MATERIAL IN FLUID PATH DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBERS 4352162, 5010440, 5016831, 5016834, 5035833 AND 5041302. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
IT WAS REPORTED THAT THE BD SYRINGE 10ML LL BNS HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL #301029 LOT #UNKNOWN. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. ITEM : 301029. LOT NUMBER: COULD NOT BE DETERMINED. ISSUE: A LARGE BROWN UNIDENTIFIED PARTICLE WAS FOUND IN ONE SYRINGE. PHOTO ATTACHED: YES- 1. SAMPLES AVAILABLE : YES - 1. THE FOLLOWING LOT NUMBERS WERE USED: ¿ 4352162 QTY. (B)(4). ¿ 5010440 QTY. (B)(4). ¿ 5016831 QTY. (B)(4). ¿ 5016834 QTY. (B)(4). ¿ 5035833 QTY. (B)(4). ¿ 5041302 QTY. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 861470 | BD SYRINGE 10ML LL BNS | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | UNKNOWN |
Patients