FDA Adverse Event Malfunction Summary report: N

MINIATURE PIN & LIGATURE CUTTER

MDR report key: 2264973 · Received August 17, 2011

Report

Report Number
2086211-2011-00005
Event Type
Malfunction
Date Received
August 17, 2011
Report Date
July 18, 2011
Manufacturer
GAC INTERNATIONAL
Product Code
EJB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED FOR EVAL AND WAS FOUND TO HAVE EXCEEDED THE EXPECTED LIFE OF THE DEVICE. WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT REC'D IN PAST TWO YRS INVOLVING A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803.

Description of Event or Problem · 1

IN THIS EVENT IT WAS REPORTED THAT THE TIP OF A MINIATURE PIN AND LIGATURE CUTTER SEPARATED OUTSIDE OF A PT'S MOUTH; THERE IS NO INDICATION THAT INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINIATURE PIN & LIGATURE CUTTER EJB GAC INTERNATIONAL 03-43

Patients

Seq Age Sex Outcome Treatment
1 LIG WIRE (SMALL)| CHAIN LIGS