FDA Adverse Event Malfunction Summary report: N

2.25MM PEDICLE PROBE

MDR report key: 22648060 · Received July 29, 2025

Report

Report Number
1723170-2025-02816
Event Type
Malfunction
Date Received
July 29, 2025
Date of Event
May 28, 2025
Report Date
July 29, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
UDI-DI
00763000121198
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H3, H6: ANALYSIS WAS PERFORMED FOR PRODUCT: 9733457, LOT NUMBER: 240506. IT WAS REPORTED THAT THE TIP WAS BENT / TWISTED. CODES B01, C07 AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THERE WAS DEFORMATION OF THE TIP OF THE CERVICAL PROBE. IT WAS NOTED THAT THE PROBABLE CAUSE OF THE ISSUE WAS THAT THE DEVICE WAS USED IN A HARD PART OF THE BONE AND BENT. THE OPERATION WAS COMPLETED BY USING A DIFFERENT INSTRUMENT. THIS ISSUE CAUSED NO SURGICAL DELAY. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1160840 2.25MM PEDICLE PROBE NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. 9733457 240506 00763000121198

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown